FDA Adverse Event Malfunction Summary report: N

ENDOPATH** XCEL* TROCAR

MDR report key: 2170482 · Received July 21, 2011

Report

Report Number
3005075853-2011-02924
Event Type
Malfunction
Date Received
July 21, 2011
Date of Event
July 12, 2011
Report Date
July 13, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS FUNCTIONALLY TESTED TO DETECT ANY LEAKING ISSUES. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK DURING INSERTION AND REMOVAL OF THE TEST PROBE THROUGH THE DEVICE. IT SHOULD BE NOTED THAT AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE REPORTED INSUFFLATIONS ISSUES. AS THE OBTURATOR WAS NOT RECEIVED AND IT IS THE PART THAT HAS THE BATCH STAMPED, WE DID NOT RECEIVE A BATCH OR LOT NUMBER FOR THE PRODUCT INVOLVED IN THIS COMPLAINT. THEREFORE, WE WERE UNABLE TO CHECK MANUFACTURING RECORDS FOR ANY RELATED NON-CONFORMANCES.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED HAND ASSISTED LOW ANTERIOR RESECTION PROCEDURE, THE SURGEON WAS USING A DEVICE AND THEY WERE ENCOUNTERING INSUFFLATIONS ISSUES; THEY REPLACED WITH IT WITH THE ANOTHER AND AFTER THE INSTRUMENT WAS INSERTED THE PRESSURE DROPPED TO 7L AND THERE WERE HISSING SOUNDS AND THE SURGEON'S VISIBILITY WAS IMPAIRED. THEY WERE USING THE RF DEVICE AND THE ENDOSCOPIC STAPLER WITH THE TROCARS. THEY WERE ABLE TO COMPLETE THE PROCEDURE WITH THE DEVICE BUT IT WAS A NUISANCE TO THE SURGEON. THERE WAS NO PATIENT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** XCEL* TROCAR LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 RF DEVICE, ENDOSCOPIC STAPLER