ECHELON* 60 ENDOPATH**
Report
- Report Number
- 3005075853-2011-02906
- Event Type
- Malfunction
- Date Received
- July 20, 2011
- Date of Event
- June 22, 2011
- Report Date
- June 22, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K051002
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INSTRUMENT B: BATCH # (B)(4), EXP DATE: 12/19/2015; 01/19/2011. THE ANALYSIS RESULTS FOUND THAT LONG60 DEVICE A WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD PRESENT. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE 3-STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. THE LONG60 INSTRUMENT B WAS RECEIVED WITH NO VISUAL NON-CONFORMANCES. THE DEVICE WAS LOADED WITH AN UNFIRED RELOAD. UPON FURTHER INSPECTION OF THE RELOAD, THE ONE-PIECE SLED WAS FOUND DAMAGED. IMPROPER LOADING OF THE RELOAD MAY DAMAGE THE SLED. IF THE RELOAD IS NOT FULLY SEATED WHEN THE DEVICE IS CLOSED, THE SLED WILL BREAK. WHEN INSERTING THE RELOAD, SLIDE IT AGAINST THE BOTTOM OF THE RELOAD JAW UNTIL THE RELOAD ALIGNMENT TAB STOPS IN THE RELOAD ALIGNMENT SLOT. SNAP THE RELOAD SECURELY IN PLACE. THE INSTRUMENT IS NOW LOADED AND READY FOR USE. PLEASE REFERENCE THE INSTRUCTION FOR USE FOR THE COMPLETE GUIDE TO LOADING AND RELOADING THE INSTRUMENT. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND ACHIEVED ITS COMPLETE 3-STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. EVENT COULD NOT BE CONFIRMED AS THE RELOAD THAT ALLEGEDLY PRODUCE THE INCOMPLETE STAPLE LINE WAS NOT RETURNED FOR ANALYSIS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC ROUX-EN-Y PROCEDURE, THE STAPLE LINE WAS NOT COMPLETE ON ONE SIDE. THE PROCEDURE WAS COMPLETED WITH A SECOND DEVICE WITH CARTRIDGES FROM A DIFFERENT LOT NUMBER. NO IMPACT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ECHELON* 60 ENDOPATH** | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |