FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 2170421 · Received July 20, 2011

Report

Report Number
6000001-2011-13240
Event Type
Malfunction
Date Received
July 20, 2011
Date of Event
June 30, 2011
Report Date
July 5, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: ONE SAMPLE CONTAINING APPROXIMATELY 120 ML OF FLUID WAS RECEIVED BY BAXTER FOR EVALUATION. HOWEVER, THE LUER WAS NOT RETURNED WITH THE DEVICE FOR EVALUATION. THE REPORTED CONDITION OF BROKEN TUBING AT THE JUNCTION OF THE LUER POST WAS CONFIRMED. BASED ON THE EVALUATION, BROKEN TUBING NEAR THE BASE OF THE STEM IS DUE TO EXCESSIVE FORCE APPLIED TO THE COMPONENT DURING FILLING. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE SAMPLE. PER REVIEW OF THE BATCH RECORDS, NO NONCONFORMANCE REPORT WAS DOCUMENTED FOR THIS LOT. ALL RELEASE CRITERIA WERE MET FOR THE BUILD OF THE LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.

Additional Manufacturer Narrative · 1

(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INTERMATE HAD A CUT IN THE TUBING BEFORE PATIENT CONNECTION. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. THE SAMPLE IS AVAILABLE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERMATE PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 11A038

Patients

Seq Age Sex Outcome Treatment
1