CONTAK RENEWAL
Report
- Report Number
- 2124215-2011-10399
- Event Type
- Malfunction
- Date Received
- July 20, 2011
- Date of Event
- May 31, 2011
- Report Date
- November 29, 2012
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. [PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
APPROXIMATELY ONE YEAR LATER THE DEVICE WAS EXPLANTED FOR REPORTED NORMAL BATTERY DEPLETION. APPROXIMATELY SIX MONTHS AFTER EXPLANT IT WAS RETURNED FOR STANDARD ANALYSIS TESTING.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE DEVICE WAS INTERROGATED PRIOR TO THE PATIENT HAVING KNEE SURGERY. UPON INTERROGATION FOUR STORED EPISODES OF NOISE WITH OVERSENSING LEADING TO A DIVERTED VENTRICULAR FIBRILLATION (VF) EVENT AND SEVERAL EPISODES OF ASYSTOLE WERE SEEN. ALL THREE EPISODES OF ASYSTOLE WERE GREATER THAN TWO SECONDS AND NOISE WAS OBSERVED ON THE SHOCK AND RIGHT VENTRICULAR CHANNELS. THE PATIENT WAS NOTED TO BE PACEMAKER DEPENDENT AND EXPERIENCING ATRIAL FIBRILLATION. BOSTON SCIENTIFIC TECHNICAL SERVICES ADVISED THE SALES REPRESENTATIVE (SR) TO DISCUSS THE SITUATION WITH THE PHYSICIAN AND EVALUATE THE SYSTEM. THE SR STATED THAT THE PATIENT WAS NOT IN A CONDITION TO PERFORM ISOMETRICS, AND THE DEVICE WAS REPROGRAMMED TO LEAST SENSITIVE WITH AN INCREASE IN THE DURATION OF 2.5 SECONDS INTO THE VF ZONE. THE PHYSICIAN OKAYED THE PATIENT FOR THE KNEE SURGERY SINCE IT HAS BEEN PREVIOUSLY DECIDED SHE WOULD BE SENT TO THE INTENSIVE CARE UNIT FOLLOWING THE PROCEDURE. NO ADDITIONAL INFORMATION WAS AVAILABLE AND NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H210 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | 4543| 1284| 4088| 4470| H210| 0185 |