VITALITY
Report
- Report Number
- 2124215-2011-09702
- Event Type
- Malfunction
- Date Received
- July 20, 2011
- Date of Event
- June 2, 2011
- Report Date
- June 2, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR (ERI) BATTERY STATUS AND WAS REPLACED WITH ANOTHER ICD. UPON RECEIPT, INITIAL LABORATORY ANALYSIS DETERMINED THAT THIS DEVICE DID NOT MEET EXPECTED LONGEVITY PREDICTIONS AS DESCRIBED IN DEVICE LABELING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T180 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | T165| 6987| E110| T180| 4016| 0175| T125| 2364L| (B)(4)| 1860| MISMATCH |