FDA Adverse Event Malfunction Summary report: N

ENDOTAK RELIANCE

MDR report key: 2170298 · Received July 20, 2011

Report

Report Number
2124215-2011-11827
Event Type
Malfunction
Date Received
July 20, 2011
Date of Event
April 12, 2011
Report Date
April 12, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
PMA / PMN Number
P910073
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

TO DATE, INFORMATION SUGGESTS THAT THIS RV LEAD REMAINS ACTIVELY IN SERVICE. IN FURTHER TROUBLESHOOTING , THE ASSOCIATED ICD WAS REMOVED FROM SERVICE. BOSTON SCIENTIFIC FINAL ANALYSIS COULD NOT CONFIRM NOR DENY THE PHYSICIAN'S CONCERN FOR SUB-PECTORAL ADVISORY RELATED DEVICE BEHAVIOR AND POSSIBLE SHORTED CONDITION. ANALYSIS OF THE RETURNED DEVICE SHOWED THE MISSING DAILY MEASUREMENTS WERE DUE TO CORRUPTED DATA BEING WRITTEN TO DEVICE MEMORY, THAT THE CAUSE OF THE CORRUPTED DATA COULD NOT BE DETERMINED, THAT INITIAL TESTING CONFIRMED ALL THERAPIES WERE AVAILABLE, AND HOW AFTER THE DEVICE FIRMWARE WAS RELOADED AND THE DEVICE PROGRAMMED TO THE SAME CLINICAL SETTINGS, NO ANOMALIES WITH THE LEAD MEASUREMENTS WERE NOTED. SUBSEQUENTLY, TESTING AGAIN CONFIRMED ALL THERAPIES WERE AVAILABLE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS TRANSVENOUS RIGHT VENTRICULAR (RV) LEAD DID EXHIBIT ERRATIC LOW SHOCK AND LOW PACE IMPEDANCE MEASUREMENTS DURING ROUTINE DEVICE FOLLOW UP, WHICH MAY BE THE RESULT OF A SHORTED LEAD CONDITION OR OTHER LEAD INTEGRITY ISSUE. THERE WAS NO NOISE PRESENT AND NO SHOCKS HAD BEEN DELIVERED. ONLY 1 ANTITACHYCARDIA PACING (ATP) WAS PRESENT. THE BOSTON SCIENTIFIC TECHNICAL SERVICES CONSULTANT RECOMMENDED FURTHER TROUBLESHOOTING SUCH AS SYNCHRONIZED 1.1 JOULES SHOCK AND SYNCHRONIZED MAXIMUM ENERGY SHOCK TO TEST THE INTEGRITY OF THE DEVICE SYSTEM AND TO ALSO CONSIDER THIS PATIENT UNPROTECTED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS BEYOND THE CONCERN FOR A SHORTED CONDITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0185

Patients

Seq Age Sex Outcome Treatment
1 37 YR Required Intervention 1861| A155| 6936| E110| 0185| 4471| (B)(4)| 6933