FDA Adverse Event Malfunction Summary report: N

VITALITY

MDR report key: 2170270 · Received July 20, 2011

Report

Report Number
2124215-2011-09595
Event Type
Malfunction
Date Received
July 20, 2011
Date of Event
June 2, 2011
Report Date
June 15, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Removal / Correction Number
Z-1620-09 TO Z-1621-09
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS IS CURRENTLY ON-GOING. UPON COMPLETION OF ANALYSIS THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED, REPLACED AND RETURNED FOR AN UNKNOWN REASON. NO FIELD ALLEGATIONS HAVE BEEN RECEIVED AGAINST THIS PRODUCT. THIS REPORT WAS CREATED DUE TO LABORATORY ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T180

Patients

Seq Age Sex Outcome Treatment
1 39 YR (B)(4)| (B)(4)