FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 2170266 · Received July 20, 2011

Report

Report Number
2124215-2011-09557
Event Type
Malfunction
Date Received
July 20, 2011
Date of Event
June 2, 2011
Report Date
June 6, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED AND IS CURRENTLY BEING ANALYZED. THIS EVENT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. NO ALLEGATIONS AGAINST THIS DEVICE WERE RECEIVED FROM THE FIELD. THIS EVENT WAS CREATED DUE TO THE INITIAL TESTING PERFORMED BY OUR POST MARKET QUALITY ASSURANCE LABORATORY. INITIAL TESTING NOTED A POSSIBLE PERFORMANCE ISSUE. NO ADVERSE PATIENT EFFECTS WERE NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T177

Patients

Seq Age Sex Outcome Treatment
1 75 YR 0155| E102| T177| 1793