VITALITY 2
Report
- Report Number
- 2124215-2011-09557
- Event Type
- Malfunction
- Date Received
- July 20, 2011
- Date of Event
- June 2, 2011
- Report Date
- June 6, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN
- Reporter Occupation
- PHYSICIAN
Narratives
THE PRODUCT HAS BEEN RECEIVED AND IS CURRENTLY BEING ANALYZED. THIS EVENT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. NO ALLEGATIONS AGAINST THIS DEVICE WERE RECEIVED FROM THE FIELD. THIS EVENT WAS CREATED DUE TO THE INITIAL TESTING PERFORMED BY OUR POST MARKET QUALITY ASSURANCE LABORATORY. INITIAL TESTING NOTED A POSSIBLE PERFORMANCE ISSUE. NO ADVERSE PATIENT EFFECTS WERE NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | 0155| E102| T177| 1793 |