FDA Adverse Event Other Summary report: N

ACRYSOF TORIC

MDR report key: 2170182 · Received June 24, 2011

Report

Report Number
1119421-2011-00794
Event Type
Other
Date Received
June 24, 2011
Date of Event
May 19, 2011
Report Date
May 27, 2011
Manufacturer
ALCON RESEARCH, LTD. / HUNTINGTON
Product Code
HQL
PMA / PMN Number
P930014
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION: THE PRODUCT WAS NOT RETURNED FOR ANALYSIS. RESULTS FROM THE PRODUCT HISTORY RECORD REVIEW INDICATED THE PRODUCT MET RELEASE CRITERIA. ROOT CAUSE: NO ROOT CAUSE COULD BE IDENTIFIED BY THE INVESTIGATION. ACTION TAKEN: INVESTIGATION HAS BEEN COMPLETED BASED ON CURRENT INFORMATION. CONCLUSION: BASED ON OUR CURRENT TRACKING, THERE ARE NO ADVERSE TRENDS FOR THIS REPORTED COMPLAINT AND BASED ON THESE FINDINGS THE RESIDUAL RISK REMAINS UNCHANGED AND AT AN ACCEPTABLE LEVEL. THERE HAVE BEEN NO OTHER COMPLAINTS REPORTED IN THE LOT NUMBER. ADDITIONAL INFORMATION HAS BEEN REQUESTED. (B)(4).

Description of Event or Problem · 1

A SURGEON REPORTED HAVING A PT WITH AN UNEXPECTED POSTOPERATIVE REFRACTION FOLLOWING INTRAOCULAR LENS (IOL) IMPLANT SURGERY. THE LENS WAS OFF BY FIVE DEGREES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACRYSOF TORIC INTRAOCULAR LENS HQL ALCON RESEARCH, LTD. / HUNTINGTON SN60T4 12007206

Patients

Seq Age Sex Outcome Treatment
1 Other