FDA Adverse Event Malfunction Summary report: N

NEXUS® BONESCALPEL® 20MM, BLUNT BLADE + IRRIGATION TUBING KIT

MDR report key: 21701612 · Received March 26, 2025

Report

Report Number
2435119-2025-00021
Event Type
Malfunction
Date Received
March 26, 2025
Date of Event
February 24, 2025
Report Date
March 26, 2025
Manufacturer
MISONIX, INC.
Product Code
LFL
PMA / PMN Number
K231117
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

NEXUS® BONESCALPEL® 20MM, BLUNT BLADE + IRRIGATION TUBING KIT (PART NUMBER 110-31-1120, LOT NUMBER 243396). ON FEBRUARY 26, 2025, MISONIX® LLC., A BIOVENTUS® CO., RECEIVED A REPORT OF AN EVENT INVOLVING A NEXUS® BONESCALPEL® 20MM, BLUNT BLADE + IRRIGATION TUBING KIT (PART NUMBER 110-31-1120, LOT NUMBER 243396) THAT OCCURRED DURING A DECOMPRESSION PROCEDURE ON (B)(6) 2025. SPECIFICALLY, THE REPORTER INDICATED THAT THE "BLADE WAS IN USE FOR A FEW MINUTES, THEN STOPPED CUTTING AND IRRIGATION WAS SPUTTERING. BLADE IS SNAPPED IN 1 SPOT." ALL PIECES OF THE BROKEN BLADE WERE REMOVED FROM THE STERILE FIELD WITHOUT ADDITIONAL MEDICAL INTERVENTION. A DELAY IN TREATMENT OF TWO MINUTES WAS REPORTED. A SERIOUS INJURY TO THE PATIENT OR USER WAS NOT REPORTED. THE DEVICE HISTORY RECORD WAS REVIEWED FOR THE NEXUS® BONESCALPEL® 20MM, BLUNT BLADE + IRRIGATION TUBING KIT (PART NUMBER 110-31-1120, LOT NUMBER 243396) IN USE AT THE TIME OF THE EVENT. THERE WERE NO DEVIATIONS FOUND DURING IN-PROCESS OR FINAL INSPECTION OF THE DEVICE. INSPECTION AND TEST RESULTS MET MISONIX SPECIFICATIONS PRIOR TO RELEASE TO COMMERCE. THE NEXUS® BONESCALPEL® 20MM, BLUNT BLADE + IRRIGATION TUBING KIT (PART NUMBER 110-31-1120) IS A STERILE KIT CONTAINING MULTIPLE COMPONENT PARTS. THE HORN (PART NUMBER E4002ABC06) AND BLADE (PART NUMBER E4003ABC20) WERE RECEIVED ASSEMBLED FOR EVALUATION. VARIOUS POINTS OF DAMAGE WERE VISIBLE ON THE BLADE AND HORN. THE INITIAL VISUAL EVALUATION CONFIRMED A FRACTURE WAS FOUND AT THE TAPER, BREAKING ALONG THE IRRIGATION TUNNEL. THE BROKEN SIDE OF THE BLADE WAS BENT OUTWARD LIKELY DUE TO A LATERAL FORCE. POWER AND IRRIGATION WERE RUN THROUGH THE RETURNED ASSEMBLED PARTS. THE ASSEMBLY MISTED MOMENTARILY AND THEN STARTED SPUTTERING IRRIGATION. AFTER APPROXIMATELY 5 SECONDS, THE BLADE FULLY FRACTURED OFF OF THE ASSEMBLY. THE HORN WAS FOUND TO STILL BE FUNCTIONAL WITH A DIFFERENT 20MM BLADE. A REVIEW OF POST MARKET SURVEILLANCE DATA FOR THE NEXUS® SYSTEM DID NOT SHOW ANY SIGNIFICANT ADVERSE TRENDS FOR BREAKAGE OF ULTRASONIC PROBE TIPS. THE CURRENT FREQUENCY OF OCCURRENCE IS WITHIN THE FREQUENCY IN THE ORIGINAL RISK MANAGEMENT REPORT. THERE IS NO CHANGE TO THE RESIDUAL RISK OR RISK-BENEFIT RATIO OF THE DEVICE. THE RISK OF HARM IS MITIGATED BY THE NATURE IN WHICH TITANIUM ULTRASONIC BLADES BREAK DURING USE. TITANIUM DOES NOT SHATTER, SPLINTER, OR CREATE MULTIPLE FRAGMENTS THAT WOULD BE DIFFICULT TO LOCATE OR REMOVE FROM THE SURGICAL FIELD. IN MOST CASES, THE BROKEN PIECES CAN BE EASILY IDENTIFIED BY DIRECT VISUAL EXAMINATION AND REMOVED FROM THE SURGICAL FIELD. IN CASES WHERE BROKEN PIECES CANNOT BE IDENTIFIED VISUALLY, ENHANCED VISUALIZATION THROUGH THE ROUTINE USE OF LOUPE-FITTED EYEGLASSES OR MICROSCOPES CAN AID IN IDENTIFYING AND REMOVING BROKEN PIECES FROM THE SURGICAL FIELD. SURGICAL SUITES TYPICALLY HAVE ACCESS TO DIAGNOSTIC IMAGING EQUIPMENT, SUCH AS X-RAY OR FLUOROSCOPY, AND WOULD BE ABLE TO QUICKLY IDENTIFY AND LOCATE ANY FRAGMENTS NOT FOUND BY DIRECT OR ENHANCED VISUALIZATION. THE BROKEN PIECES THAT REMAIN IN THE OPERATIVE FIELD CAN THEREFORE BE FOUND AND EXTRACTED QUICKLY AND EASILY WITHOUT SIGNIFICANT DELAY IN THE SURGICAL PROCEDURE. THE INSTRUCTIONS FOR USE MANUAL (100-10-1000, REVISION L) FOR THE NEXUS® ULTRASONIC SURGICAL ASPIRATION SYSTEM CONTAINS THE FOLLOWING WARNINGS AND CAUTIONS TO PREVENT BLADE BREAKAGE. WARNING: THE NEXUS® ULTRASONIC SURGICAL ASPIRATOR SYSTEM IS INTENDED TO BE USED IN VARIOUS TYPES OF INVASIVE, SURGICAL PROCEDURES. THERE MAY BE INDIRECT DANGER TO THE PATIENT SHOULD THE DEVICE FAIL DURING THE PROCEDURE. IT IS RECOMMENDED THAT THE FACILITY FOLLOWS ITS BACK-UP EQUIPMENT PROTOCOLS. WARNING: ULTRASONIC PROBE TIPS CAN BREAK UNDER EXCESSIVE USE IN EXTREME CONDITIONS, E.G. WHEN CUTTING FOR EXTENDED DURATION IN TIGHT CAVITIES WITH LIMITED LATERAL MOTION. THE PROBE TIP COULD BREAK INTO TWO OR MORE FRAGMENTS WITH THE MAIN FRAGMENT REMAINING ATTACHED TO THE HANDPIECE. ALL FRAGMENTS MUST BE RETRIEVED IMMEDIATELY FROM THE SURGICAL SITE. THE FRAGMENTS SHOULD BE CHECKED TO ENSURE THAT NO FURTHER PIECES ARE MISSING. IT IS POSSIBLE THAT A FRAGMENT IS PROPELLED OUTSIDE OF THE SURGICAL CAVITY. DIAGNOSTIC IMAGING, SUCH AS X-RAY, MUST BE USED IF A FRAGMENT CANNOT BE FOUND TO CONFIRM THAT THE BROKEN PIECE IS OUTSIDE OF THE SURGICAL CAVITY. WARNING: BREAKAGE OF ULTRASONIC PROBE TIPS WILL RESULT IN SHARP EDGES THAT CAN BE HARMFUL TO SOFT TISSUE EVEN WITHOUT ACTIVATION OF ULTRASOUND. PROBE TIPS CAN BEND OR DEFORM BEFORE THEY ACTUALLY BREAK. PROBE TIPS SHOWING SIGNS OF DEFORMATION OR CRACKING SHOULD BE REPLACED IMMEDIATELY SINCE PROBE TIP BREAKAGE IS OTHERWISE IMMINENT. DO NOT BEND OR TWIST THE ULTRASONIC PROBE TIPS SINCE IT REDUCES THE STRUCTURAL INTEGRITY AND CAN RESULT IN PROBE TIP BREAKAGE DURING USE. DISPOSE OF DEFORMED OR BROKEN PROBE TIPS IMMEDIATELY IN A SHARPS CONTAINER IN ACCORDANCE WITH YOUR FACILITY BIOLOGICAL HAZARDOUS WASTE PROCEDURE. WARNING: DURING SYSTEM CHECK, MAKE SURE THE PROBE TIP OF THE HANDPIECE IS FREE FROM CONTACT WITH ANY OBJECT. ALLOWING CONTACT WITH THE PROBE TIP MAY RESULT IN DAMAGE AND/OR PERSONAL INJURY. CAUTION THE SYSTEM CHECK SHOULD ALWAYS BE DONE IN ADVANCE OF PREPARING PATIENT FOR SURGERY TO MINIMIZE RISK TO PATIENT IN CASE OF SYSTEM MALFUNCTION. REFER TO SECTION 7.11 - SYSTEM CHECK. CAUTION: LOOSE PROBE TIP/TISSUE CONTACT UPON AN INITIAL BONE INCISION CAN CAUSE A THIN PROBE TIP TO RESONATE NOT ONLY LONGITUDINALLY BUT ALSO TRANSVERSELY. THIS CAN CAUSE A THIN PROBE TIP TO BREAK. IT IS NECESSARY TO ENGAGE BONE ACTIVELY AND WITH A MINIMAL PROBE TIP PRESSURE GREATER THAN ZERO IN ORDER TO PREVENT THE SHATTERING. CAUTION: CONTACT OF THE ULTRASONIC PROBE TIP OR THE EXPOSED EXTENSION WITH METAL, SURGICAL INSTRUMENTS OR OTHER OBJECTS DURING ULTRASOUND USE MUST BE AVOIDED. SUCH CONTACT CAN DAMAGE THE ULTRASONIC COMPONENTS VERY EASILY AND MAY RESULT IN COMPROMISED PERFORMANCE, INCLUDING FAILURE. DISCARD ANY EXTENSIONS OR PROBE TIPS THAT SHOW SIGNS OF DAMAGES LIKE GOUGES, NICKS OR FRACTURES. EXTERNAL ASPIRATION MAY BE USED BUT IT IS RECOMMENDED THAT A PLASTIC SUCTION PROBE TIP SHOULD BE USED WHEN IN PROXIMITY WITH THE PROBE TIP. CAUTION: THE NEXUS® SYSTEM SHOULD BE FULLY TESTED AND INSPECTED PRIOR TO EACH PROCEDURE. THE CONSOLE, FOOTSWITCH, HANDPIECES, ALL CABLES AND ACCESSORIES SHOULD BE EXAMINED FOR PROPER APPEARANCE AND CONDITION. THE INVESTIGATION HAS BEEN CONCLUDED.

Description of Event or Problem · 0

NEXUS® BONESCALPEL® 20MM, BLUNT BLADE + IRRIGATION TUBING KIT (PART NUMBER 110-31-1120, LOT NUMBER 243396). ON FEBRUARY 26, 2025, MISONIX® LLC., A BIOVENTUS® CO., RECEIVED A REPORT OF AN EVENT INVOLVING A NEXUS® BONESCALPEL® 20MM, BLUNT BLADE + IRRIGATION TUBING KIT (PART NUMBER 110-31-1120, LOT NUMBER 243396) THAT OCCURRED DURING A DECOMPRESSION PROCEDURE ON (B)(6) 2025. SPECIFICALLY, THE REPORTER INDICATED THAT THE "BLADE WAS IN USE FOR A FEW MINUTES, THEN STOPPED CUTTING AND IRRIGATION WAS SPUTTERING. BLADE IS SNAPPED IN 1 SPOT." ALL PIECES OF THE BROKEN BLADE WERE REMOVED FROM THE STERILE FIELD WITHOUT ADDITIONAL MEDICAL INTERVENTION. A DELAY IN TREATMENT OF TWO MINUTES WAS REPORTED. A SERIOUS INJURY TO THE PATIENT OR USER WAS NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
570787 NEXUS® BONESCALPEL® 20MM, BLUNT BLADE + IRRIGATION TUBING KIT 20MM, BLUNT BLADE + IRRIGATION TUBING KIT LFL MISONIX, INC. 110-31-1120 243396

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other