FDA Adverse Event Malfunction Summary report: N

COBAS E 411 ANALYZER (DISK SYSTEM)

MDR report key: 21700977 · Received March 26, 2025

Report

Report Number
1823260-2025-00895
Event Type
Malfunction
Date Received
March 26, 2025
Date of Event
February 21, 2025
Report Date
April 25, 2025
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
UDI-DI
04015630937103
PMA / PMN Number
K961481
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REAGENT LOT NUMBER IS 802253. THE EXPIRATION DATE WAS NOT PROVIDED. THE FIELD SERVICE REPRESENTATIVE FOUND AN ISSUE WITH THE MEASURING CELL CARRYOVER. HE PERFORMED MAINTENANCE AND REPLACED AN OVERDUE MEASURING CELL, SIPPER PROBE, AND PINCH VALVE TUBING. HE NOTED THAT THE INSTRUMENT WAS VERY DIRTY. HE THOROUGHLY CLEANED THE INSTRUMENT. HE PERFORMED ADJUSTMENTS, CHECKS, AND TESTS WITH ACCEPTABLE RESULTS. THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative · 0

THE CALIBRATION DATA INDICATED SIGNALS FOR ONE CALIBRATOR LEVEL THAT WERE HIGHER THAN EXPECTED. THE QC WAS ACCEPTABLE. A GENERAL REAGENT ISSUE WAS EXCLUDED. AFTER SERVICE, NO ISSUES WERE REPORTED BY THE CUSTOMER. THE INVESTIGATION DETERMINED THE SERVICE ACTIONS RESOLVED THE ISSUE.

Description of Event or Problem · 0

THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS TROPONIN T GEN 5 STAT RESULTS FOR 3 PATIENT SAMPLES ON A COBAS E 411 ANALYZER (DISK SYSTEM). PATIENT (B)(6): THE INITIAL RESULT WAS 78 NG/L, AND THE REPEATED RESULT WAS <6.00 NG/L. PATIENT (B)(6): ON (B)(6) 2025, THE INITIAL RESULT WAS 77 NG/L, AND THE REPEATED RESULT WAS <6.00 NG/L. PATIENT (B)(6): ON (B)(6) 2025, THE INITIAL RESULT WAS 50.27 NG/L, AND THE REPEATED RESULT WAS <6.00 NG/L. THE SAMPLES WERE REPEATED BECAUSE THE INITIAL RESULTS WERE QUESTIONED BY THE PHYSICIAN. THE REPEATED RESULTS WERE OBTAINED ON ANOTHER MODULE AND WERE BELIEVED TO BE CORRECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
785956 COBAS E 411 ANALYZER (DISK SYSTEM) IMMUNOCHEMISTRY ANALYZER JJE ROCHE DIAGNOSTICS 04015630937103

Patients

Seq Age Sex Outcome Treatment
1 65 YR Female