FDA Adverse Event Injury Summary report: N

AMS 700 CX MS PUMP

MDR report key: 21699425 · Received March 26, 2025

Report

Report Number
2124215-2025-18068
Event Type
Injury
Date Received
March 26, 2025
Date of Event
March 11, 2025
Report Date
May 28, 2025
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
JCW
UDI-DI
00878953003504
PMA / PMN Number
N970012
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

UPON RECEIPT OF THIS INFLATABLE PENILE PROSTHESIS AT OUR QUALITY ASSURANCE LABORATORY, THIS DEVICE WAS THOROUGHLY ANALYZED. MICROSCOPIC ANALYSIS IDENTIFIED BOTH CYLINDERS HAD WEAR AT FOLD IN THE PROXIMAL END, MIDDLE, AND DISTAL END OF THE CYLINDER AND, NO DAMAGE OR ABNORMALITIES WERE FOUND ON PUMP. THE DEVICE WAS TESTED TOO, CONCLUDING PUMP AND BOTH CYLINDERS PASSED LEAKAGE TEST; THE PUMP PASSED THE FUNCTIONAL TEST TOO. THE RESERVOIR WAS ALSO ANALYZED, IDENTIFYING IT HAD WEAR AT A FOLD IN RESERVOIR SHELL AND PASSING THE LEAKAGE TEST. THE ALLEGATION OF AIR IN THE SYSTEM/COMPONENT WAS NOT CONFIRMED. BASED ON THE INFORMATION PROVIDED, CAUSE NOT ESTABLISHED WAS SELECTED AS THE MOST PROBABLE CAUSE FOR THE COMPLAINT. CONCOMITANT PRODUCT UDI FOR RESERVOIR IS (B)(4). CONCOMITANT PRODUCT UDI FOR LGX PRECONNECT IS (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE INFLATABLE PENILE PROSTHESIS (IPP) PUMP HAS AIR POCKET AND DOES NOT WORK. A SURGICAL PROCEDURE WAS PERFORMED TO REPLACE THE SYSTEM. THERE WERE NO FURTHER PATIENT COMPLICATIONS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE INFLATABLE PENILE PROSTHESIS (IPP) PUMP HAS AIR POCKET AND DOES NOT WORK. A SURGICAL PROCEDURE WAS PERFORMED TO REPLACE THE SYSTEM. THERE WERE NO FURTHER PATIENT COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
716185 AMS 700 CX MS PUMP PROSTHESIS, PENIS, INFLATABLE JCW BOSTON SCIENTIFIC CORPORATION 72404237 1100195243 00878953003504

Patients

Seq Age Sex Outcome Treatment
1 56 YR Male Hospitalization| R LGX PRECONNECT: LOT: 1100494478, UPN: 72404252-10.| RESERVOIR: LOT: 110028199.| RESERVOIR: LOT: 1100482011, UPN: 720185-01.