FDA Adverse Event Malfunction Summary report: N

BD PHOENIX¿ NMIC-306

MDR report key: 21698688 · Received March 26, 2025

Report

Report Number
1119779-2025-00216
Event Type
Malfunction
Date Received
March 26, 2025
Date of Event
February 22, 2025
Report Date
May 27, 2025
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
LON
UDI-DI
30382904492923
PMA / PMN Number
SEE H11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: K020322, K022129, K023444, K023634, K023858, K024153, K031530, K031699, K031912, K032299, K032567, K032655, K033362, K033560, K041384, K042932, K052269, K060214, K060217, K060257, K060444, K060447, K061327, K061355, K062207, K062944, K063301, K063486, K063573, K063811, K063824, K071623, K132674, K132909, K151320, K181665, K173252, K190905, K163637, K173523, K033458, AND K123266. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: THIS COMPLAINT IS FOR HIGH MIC ERTAPENEM (ETP) WITH ENTEROBACTER CLOACAE COMPLEX WHEN USING PHOENIX PANEL NMIC-306 (CATALOG NUMBER 449292) BATCH NUMBER 4353531. THE CUSTOMER DID NOT PROVIDE PANEL RETURNS OR ISOLATES BUT PROVIDED PHOENIX GENERATED LAB REPORTS FOR THE INVESTIGATION. THE LAB REPORTS SHOW HIGH MIC RESULTS FOR ETP WITH ENTEROBACTER CLOACAE COMPLEX WHEN USING THE COMPLAINT BATCH. THE LAB REPORTS ALSO NOTE ACINETOBACTER BAUMANII/CALCOACITECUS COMPLEX WITH INTRINSIC RESISTANCE TO ETP. TO INVESTIGATE, RETENTION PANELS OF THE COMPLAINT BATCH WERE INOCULATED WITH IN HOUSE ISOLATE ENTEROBACTER CLOACAE COMPLEX ENF 11061 AND PLACED IN A PHOENIX M50 TO EVALUATE FOR ETP MIC RESULTS. NEXT, CONTROL PANELS OF THE SAME MATERIAL BUT DIFFERENT BATCH WERE INOCULATED WITH IN HOUSE ISOLATE ENTEROBACTER CLOACAE COMPLEX ENF 11061 AND PLACED IN A PHOENIX M50 TO EVALUATE FOR ETP MIC RESULTS. AT THE END OF THE RUN, ALL PANELS RETURNED THE EXPECTED ETP MIC RESULTS. THIS COMPLAINT IS NOT CONFIRMED. THE BATCH HISTORY RECORD WAS SATISFACTORY, AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. COMPLAINT TRENDING WAS PERFORMED, AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. PLEASE CONTINUE TO COMMUNICATE ANY ADDITIONAL CONCERNS.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC-306 A PATIENT ISOLATE (ACINETOBACTER CALCOACETICUS) HAD A RESISTANT RESULT FOR THE DRUG ERTAPENEM WHILE THE OTHER DRUGS OF THE PANEL WERE SENSITIVE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 1 OF 5.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC-306 A PATIENT ISOLATE (ACINETOBACTER CALCOACETICUS) HAD A RESISTANT RESULT FOR THE DRUG ERTAPENEM WHILE THE OTHER DRUGS OF THE PANEL WERE SENSITIVE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 1 OF 5.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1059339 BD PHOENIX¿ NMIC-306 SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION LON BECTON DICKINSON & CO. (SPARKS) 4353531 30382904492923

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown