COULTER LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2011-00882
- Event Type
- Malfunction
- Date Received
- July 15, 2011
- Date of Event
- November 19, 2009
- Report Date
- November 20, 2009
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K061574
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
THIS SAMPLE WAS BEING TESTED DURING THE POST-INSTALLATION STUDY OF SOFTWARE VERSION 2D1. THE INSTRUMENT WAS WITHIN QUALITY CONTROL SPECS WITH RESPECT TO CONTROLS (ACCURACY) AND REPRODUCIBILITY (PRECISION). THE SENSITIVITY FOR FLAGGING WAS SET AT [3201], WHICH IS HIGH-LEVEL FOR BLASTS, MID-LEVEL FOR VARIANT LYMPHOCYTES AND SET TOO OFF FOR IMM. NE 1 AND LOW LEVEL FOR IMM. NE 2 FLAGGING. NOTE: IMM NE 1 IS A FLAG FOR SUSPECT IMMATURE NEUTROPHILS, PRIMARILY BANDS. IMM NE 2 IS A FLAG FOR SUSPECT IMMATURE NEUTROPHILS, PRIMARILY METAMYELOCYTES, MYELOCYTES, AND PROMYELOCYTES. SERVICE WAS NOT DISPATCHED FOR THIS EVENT. RAW DATA ANALYSIS DETERMINED THAT THE DATA FEATURES FROM THE SAMPLE WERE ON THE BOUNDARY CONDITION OF THE BLAST FLAGGING RULE AND DID GENERATE IMM NE2 FLAG. THE ROOT CAUSE FOR THE LACK OF FLAG FOR BLAST CELLS FOR THIS PARTICULAR PT WAS ATTRIBUTED TO THE SPECIMEN. THE SPECIMEN WAS NOT SIGNIFICANTLY ABNORMAL TO TRIGGER BLAST FLAGGING. HOWEVER, THE ANALYZER DID GENERATE AN IMM. NE 2 DIFFERENTIAL FLAG THAT ALERTS THE OPERATOR TO FURTHER REVIEW THE PT SAMPLE. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW CONDUCTED BETWEEN (B)(4), 2008 AND (B)(4), 2010 OF COMPLAINTS FOR ADD'L REPORTABLE EVENTS.
CUSTOMER REPORTED ERRONEOUS DIFFERENTIAL RESULTS WERE OBTAINED FOR ONE PT SAMPLE WHEN USING A COULTER LH 750 HEMATOLOGY ANALYZER. THE TEST RESULTS WERE DETERMINED TO BE ERRONEOUS WHEN COMPARED TO A MANUAL BLOOD SMEAR DIFFERENTIAL WITH BLAST CELLS. THERE WERE NO INSTRUMENT GENERATED FLAGS FOR THE PRESENCE OF BLAST CELLS. THERE WAS AN INSTRUMENT GENERATED IMM NE 2 DIFFERENTIAL FLAG. IMM NE 2 IS A FLAG FOR SUSPECT IMMATURE NEUTROPHILS, PRIMARILY METAMYELOCYTES, MYELOCYTES, AND PROMYELOCYTES. ERRONEOUS TEST RESULTS WERE NOT REPORTED OUT OF THE LAB. NO REPORTS OF DEATH OR SERIOUS INJURY AND NO EFFECT TO PT TREATMENT AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER LH 750 HEMATOLOGY ANALYZER | GKZ | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |