Report Number
3015543569-2025-00002
Event Type
Injury
Date Received
March 26, 2025
Report Date
September 10, 2025
Manufacturer
EMBODY, INC
Product Code
OWY
UDI-DI
00810063820053
PMA / PMN Number
K201572
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: A2, A3, A4, B4, B5, B7, D6A, G2, G3, G6, H1, H2, H6, H11 CORRECTED SECTIONS: D1, D2,D4, E1, E2, E3 THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4, B5, D4, G3, G6, H2, H3, H4, H6, H11 NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING RELATED TO THE REPORTED EVENT. MEDICAL RECORDS WERE NOT PROVIDED. CRF WAS REVIEWED BY A HEALTH CARE PROFESSIONAL. REVIEW OF THE CRF IDENTIFIED THE FOLLOWING: RIGHT HIP PAIN, TROCHANTERIC BURSITIS WITH TREATMENT OF STEROID INJECTION A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. NO CORRECTIVE ACTIONS, PREVENTIVE ACTIONS, OR FIELD ACTIONS RESULTED AFTER INVESTIGATION OF THIS EVENT. THE REPORTED EVENT IS UNCONFIRMED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

(B)(4). CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION, AS THE DEVICE REMAINS IMPLANTED. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT EXPERIENCED RIGHT HIP PAIN, TROCHANTERIC BURSITIS, AND RECEIVED A STEROID INJECTION POST INITIAL IMPLANTATION. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT HAD AN UNKNOWN PROCEDURE AND EXPERIENCED RIGHT HIP PAIN, TROCHANTERIC BURSITIS, AND RECEIVED A STEROID INJECTION. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1131454 TAPESTRY 40X30MM MESH, SURGICAL OWY EMBODY, INC T-250 00810063820053

Patients

Seq Age Sex Outcome Treatment
1 63 YR Male Other