Report Number
2182208-2025-01499
Event Type
Injury
Date Received
March 25, 2025
Date of Event
January 1, 2024
Report Date
March 25, 2025
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS INFORMATION IS BASED ENTIRELY ON JOURNAL LITERATURE. MEDTRONIC WAS MADE AWARE OF THIS EVENT THROUGH A SEARCH OF LITERATURE PUBLICATIONS. OF NOTE, MULTIPLE PATIENTS AND MULTIPLE MANUFACTURERS WERE NOTED IN THE ARTICLE; HOWEVER, A ONE-TO-ONE CORRELATION COULD NOT BE MADE WITH UNIQUE PRODUCT SERIAL/LOT NUMBERS. THE BASELINE GENDER/AGE CHARACTERISTICS IS MALE/64 YEARS OLD. THE MODEL LISTED IN THE REPORT IS A REPRESENTATIVE OF THE MODEL FAMILY (CAPSURE AND VITATRON), AS THERE IS NO SPECIFIC MODEL LISTED. WITHOUT A LOT NUMBER OR DEVICE SERIAL NUMBER, THE MANUFACTURING DATE CANNOT BE DETERMINED. SINCE NO DEVICE ID WAS PROVIDED, IT IS UNKNOWN IF THIS EVENT HAS BEEN PREVIOUSLY REPORTED. A REQUEST FOR ADDITIONAL INFORMATION WAS MADE. IF ADDITIONAL INFORMATION IS OBTAINED REGARDING THIS EVENT, IT WILL BE ADDED, AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY. REFERENCED ARTICLE: ACCURATE DETECTION OF LEAD MALFUNCTION FROM ECG-DERIVED BIPOLAR PACING STIMULUS AMPLITUDE. HEART RHYTHM. 2024; 21:2226¿2231. DOI: 10.1016/J.HRTHM.2024.03.1814. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

LITERATURE WAS REVIEWED REGARDING DETECTION OF LEAD MALFUNCTION. THE AUTHORS DESCRIBED TRANSVENOUS LEAD MALFUNCTIONS IN WHICH THE LEADS EXHIBITED EVIDENCE OF INSULATION BREAKDOWN, LEAD NOISE, MYOPOTENTIAL OVERSENSING, OR A SIGNIFICANT DROP IN LEAD IMPEDANCE. THE LEADS WERE REVISED. THE STATUS OF THE LEADS IS UNKNOWN. NO ADVERSE PATIENT EFFECTS OR ADDITIONAL PRODUCT PERFORMANCE ISSUES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1719355 MEDTRONIC LEAD IMPLANTABLE PACEMAKER PULSE-GENERATOR DXY MEDTRONIC, INC. MDT-LEAD

Patients

Seq Age Sex Outcome Treatment
1 64 YR Male Hospitalization| R UNKNOWN COMPETITOR CIED