FDA Adverse Event Malfunction Summary report: N

COMPOUNDERS

MDR report key: 2167745 · Received July 19, 2011

Report

Report Number
6000001-2011-13020
Event Type
Malfunction
Date Received
July 19, 2011
Date of Event
June 15, 2011
Report Date
June 15, 2011
Manufacturer
BAXTER HEALTHCARE - BUFFALO GROVE
Product Code
NEP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE WAS DETERMINED TO BE A DAMAGED CERAMIC FUSE. TO CORRECT THE CONDITION, THE CERAMIC FUSE WAS REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT. THIS DEVICE IS CLASS II EXEMPT FROM HAVING A 510(K) NUMBER. ITS ADDITIONAL 510(K) NUMBER IS K955622.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED. THIS DEVICE IS EXEMPT FROM HAVING A 510(K) NUMBER. HOWEVER, THIS PRODUCT IS BEING REPORTED BECAUSE IT IS SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE U.S. ITS ADDITIONAL 510(K) NUMBER IS K955622.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, AN AUTOMIX COMPOUNDER REQUIRED REPLACEMENT OF A 302A 250V 1/4 X 1 11/4 CERAMIC SLO BLOW FUSE. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.