FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 21661582 · Received March 21, 2025

Report

Report Number
9617229-2025-04678
Event Type
Injury
Date Received
March 21, 2025
Date of Event
October 1, 2023
Report Date
August 28, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: D9 H3 H6. LABORATORY ANALYSIS SUMMARY: THE DEVICE RELATED TO THE REPORTED EVENTS OF RUPTURE WAS RECEIVED ON JULY 17, 2025, WITH LOT NUMBER 2933929. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINT CODES ARE: ¿ RUPTURE: OBSERVED AN OPENING ASSESSED AS FOLD FLAW OPENING. AS PER THE INVESTIGATION PROCEDURE, CREASES, WEAR ABRASION AND NON-PENETRATING NICK WERE OBSERVED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. DEVICE WAS EXPLANTED AND REPLACED WITH A NON-ABBVIE DEVICE.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. DEVICE WAS EXPLANTED AND REPLACED WITH A NON-ABBVIE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1082979 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2933929

Patients

Seq Age Sex Outcome Treatment
1 62 YR Female Required Intervention