PERCLOSE¿ PROSTYLE¿
Report
- Report Number
- 2024168-2025-02928
- Event Type
- Injury
- Date Received
- March 20, 2025
- Date of Event
- February 25, 2025
- Report Date
- April 14, 2025
- Manufacturer
- ABBOTT VASCULAR INC.
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.
ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS NOT CONFIRMED AS NOT ALL COMPONENTS WERE RETURNED FOR ANALYSIS. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. BASED ON THE REPORTED INFORMATION AND OBSERVATIONS FROM THE RETURNED ANALYSIS, THE INVESTIGATION DETERMINED THAT THE REPORTED ISSUE AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO OPERATIONAL CONTEXT. IT IS LIKELY THAT AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D4: LOT # CORRECTION FROM 4101241 TO 5011641. D4: EXPIRATION DATE CORRECTION FROM 9/30/2026 TO 12/31/2026. D4: PRIMARY UDI NUMBER CORRECTION FROM (B)(4). H4: CORRECTION DEVICE MFG DATE FROM 10/12/2024 TO 1/16/2025.
IT WAS REPORTED THAT THIS WAS A VENOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL VEIN USING A PROSTYLE AFTER A FEMORAL ABLATION USING AN 8F SHEATH. REPORTEDLY, WHEN THE PLUNGER WAS REMOVED, THERE WAS NO SUTURE ATTACHED. AN ADDITIONAL PROSTYLE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 453959 | PERCLOSE¿ PROSTYLE¿ | DEVICE, HEMOSTASIS, VASCULAR | MGB | ABBOTT VASCULAR INC. | 12773-03 | 5011641 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Male | Required Intervention |