FDA Adverse Event Injury Summary report: N

PERCLOSE¿ PROSTYLE¿

MDR report key: 21657515 · Received March 20, 2025

Report

Report Number
2024168-2025-02928
Event Type
Injury
Date Received
March 20, 2025
Date of Event
February 25, 2025
Report Date
April 14, 2025
Manufacturer
ABBOTT VASCULAR INC.
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.

Additional Manufacturer Narrative · 0

ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS NOT CONFIRMED AS NOT ALL COMPONENTS WERE RETURNED FOR ANALYSIS. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. BASED ON THE REPORTED INFORMATION AND OBSERVATIONS FROM THE RETURNED ANALYSIS, THE INVESTIGATION DETERMINED THAT THE REPORTED ISSUE AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO OPERATIONAL CONTEXT. IT IS LIKELY THAT AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D4: LOT # CORRECTION FROM 4101241 TO 5011641. D4: EXPIRATION DATE CORRECTION FROM 9/30/2026 TO 12/31/2026. D4: PRIMARY UDI NUMBER CORRECTION FROM (B)(4). H4: CORRECTION DEVICE MFG DATE FROM 10/12/2024 TO 1/16/2025.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS A VENOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL VEIN USING A PROSTYLE AFTER A FEMORAL ABLATION USING AN 8F SHEATH. REPORTEDLY, WHEN THE PLUNGER WAS REMOVED, THERE WAS NO SUTURE ATTACHED. AN ADDITIONAL PROSTYLE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
453959 PERCLOSE¿ PROSTYLE¿ DEVICE, HEMOSTASIS, VASCULAR MGB ABBOTT VASCULAR INC. 12773-03 5011641

Patients

Seq Age Sex Outcome Treatment
1 68 YR Male Required Intervention