PERCLOSE¿ PROSTYLE¿
Report
- Report Number
- 2024168-2025-02915
- Event Type
- Injury
- Date Received
- March 20, 2025
- Date of Event
- March 5, 2025
- Report Date
- May 9, 2025
- Manufacturer
- ABBOTT VASCULAR INC.
- Product Code
- MGB
- UDI-DI
- 08717648235184
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.
ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS CONFIRMED. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D4: CORRECTED PRIMARY UDI NUMBER FROM (B)(4). D4: CORRECTED LOT NUMBER FROM 5011841 TO 5011641. H4: CORRECTED DEVICE MFG DATE.
IT WAS REPORTED THAT THIS WAS A VENOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL VEIN USING A PROSTYLE AFTER A FEMORAL ABLATION USING AN 8F SHEATH. REPORTEDLY, WHEN THE PLUNGER WAS REMOVED, THERE WAS NO SUTURE ATTACHED. AN ADDITIONAL PROSTYLE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 537734 | PERCLOSE¿ PROSTYLE¿ | DEVICE, HEMOSTASIS, VASCULAR | MGB | ABBOTT VASCULAR INC. | 12773-03 | 5011641 | 08717648235184 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Female | Required Intervention |