FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION¿ S8 SYSTEM

MDR report key: 21653697 · Received March 20, 2025

Report

Report Number
1723170-2025-01397
Event Type
Malfunction
Date Received
March 20, 2025
Date of Event
March 13, 2025
Report Date
April 16, 2025
Manufacturer
MEDTRONIC NAVIGATION, INC
Product Code
HAW
UDI-DI
00763000272739
PMA / PMN Number
K162309
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9 735821, H6: MULTIPLE FDD/ANNEX A CODES WERE REPORTED. A05 WAS CODED FOR LOCALIZER FAULTED MESSAGE WITH AN AMBER LIGHT, BUMP SENSOR AND INTERNAL TEMPERATURE WERE HIGHLIGHTED. A1102 WAS CODED FOR LOCALIZER FAULTED MESSAGE. THE SYSTEM WAS SERVICED IN THE FIELD BY A MEDTRONIC REPRESENTATIVE. THE CAMERA WAS REPLACED. IT WAS REPLACED. CODES B01, C08, AND D02 ARE APPLICABLE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. EDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H2, H3) CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9735821, LOT #: P903590, UBD: UNKNOWN, UDI#: UNKNOWN THE RETURNED POSITIONING SENSOR UNIT (PSU) HAD SCRATCHES ON THE HOUSING AND LENSES. A CHECK OF THE EVENT LOG SHOWED INTERMITTENT FIRMWARE INCOMPATIBILITY. THERE WAS A BATTERY VOLTAGE LOW MESSAGE ALONG WITH BUMP DETECTED AND STORAGE TEMPERATURE EXCEEDED. THE PSU FAILED AN ACCURACY TEST (AAK) AT .973MM WITH A PASSING THRESHOLD OF .250MM. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE MANUFACTURER REPRESENTATIVE (REP) RECEIVED A LOCALIZER FAULTED MESSAGE WITH AN AMBER LIGHT. DURING TROUBLESHOOTING, THE REP INDICATED THE NETWORK DEVICE INTERFACE (NDI) TOOL BOX- BUMP SENSOR AND INTERNAL TEMPERATURE WERE HIGHLIGHTED. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1212155 STEALTHSTATION¿ S8 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC 9735665 00763000272739

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown "SEE H11...."| SEE H11