SYRINGE 10ML SALINE FILL CHINA SP
Report
- Report Number
- 1911916-2025-00208
- Event Type
- Malfunction
- Date Received
- March 19, 2025
- Date of Event
- February 20, 2025
- Report Date
- March 3, 2025
- Manufacturer
- BECTON DICKINSON
- Product Code
- NGT
- UDI-DI
- 00382903065950
- PMA / PMN Number
- K161552
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DEVICE EVALUATION: IT WAS REPORTED THE FLUSHER WAS FOUND DAMAGED. TO AID IN THE INVESTIGATION, ONE PHOTO WAS RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. THE PHOTO SHOWS TWO SYRINGES OUT OF THE PACKAGING FLOW WRAP WITH THE SYRINGE BARREL FLANGE DAMAGED. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THIS CONDITION OCCURS IF THERE IS A JAM DURING THE PLUNGER ROD ASSEMBLY PROCESS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 306595, LOT 4250009. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. VERIFICATION OF THE SYRINGE PLUNGER ROD PROCESS WAS PERFORMED. THE SETTINGS AND ALIGNMENT OF THE INFEED SCROLL WERE CORRECT. THE FLOW OF PRODUCT WAS GOOD. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
ON (B)(6) 2025, THE BD PRE-FILLED CATHETER FLUSHER WAS USED TO FLUSH THE PATIENT'S CATHETER. IT WAS FOUND THAT THE FLUSHER WAS DAMAGED, WHICH COULD EASILY CAUSE THE MEDICAL STAFF'S HANDS TO BE SCRATCHED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 690374 | SYRINGE 10ML SALINE FILL CHINA SP | SALINE, VASCULAR ACCESS FLUSH | NGT | BECTON DICKINSON | 4250009 | 00382903065950 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |