FDA Adverse Event Malfunction Summary report: N

BD IAG BC PRO GLOBAL WING

MDR report key: 21640867 · Received March 19, 2025

Report

Report Number
1710034-2025-00426
Event Type
Malfunction
Date Received
March 19, 2025
Date of Event
February 12, 2025
Report Date
June 16, 2025
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903829446
PMA / PMN Number
UNK
Removal / Correction Number
MDS-25-5274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 382944 AND LOT NUMBER 4298108. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. BASED ON THE AVAILABLE INFORMATION, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION OF FA#.

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD IAG BC PRO GLOBAL WING NEEDLE WAS SLOW TO RETRACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: DESCRIPTION OF THE FACTS, CIRCUMSTANCES AND MEANS USED: THE AUTO-GUARD DEVICE DOES NOT WORK OR RETRACTS VERY SLOWLY. MEASURES TAKEN: 1 YES. FREQUENCY: 5 VERY PROBABLE (MORE THAN ONCE A WEEK). DESCRIPTION OF THE CONSEQUENCES: RISK OF NEEDLE STICK AND SPLASHING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1565322 BD IAG BC PRO GLOBAL WING PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 4298108 00382903829446

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown