AUTOSOFT 90
Report
- Report Number
- 3003442380-2025-03997
- Event Type
- Malfunction
- Date Received
- March 19, 2025
- Date of Event
- February 17, 2025
- Report Date
- March 27, 2025
- Manufacturer
- UNOMEDICAL UM-D
- Product Code
- FPA
- UDI-DI
- 05705244018129
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL (B)(4) - DEVICE 2 OF 2.
SUPPLEMENTAL REPORT 01 - (B)(4). THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED. THE BATCH 6000764 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION (WI) GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 2 AND WI GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 2 FOR THE CODE SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE. COMPLAINT INVESTIGATIONS. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. TEST RESULTS: VISUAL TEST ACCORDING TO WI VERSION 2 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL FLOW TEST 1 ACCORDING TO WI VERSION 2 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6000764 WAS MANUFACTURED ACCORDING TO THE WI VERSION 103 MANUFACTURED IN THE LINE 1, ON 26/APR/2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 21/MAR/2025 AGAINST MALFUNCTION CODE SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE AND LOT 6000764 AND NO OTHER COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT 6000764 AND MALFUNCTION CODE. CONCLUSION SUMMARY OF THE RELATED EVENT: AS A RESULT OF THE FOLLOWING: NO DEFECT ON TESTS FOR REFERENCE SAMPLES, NO HARM, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, NO OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION AND MALFUNCTION CODE. NO FURTHER ACTIONS ARE REQUIRED. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITY.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED TWO INFUSION SET CANNULA KINKED EVENT ON 17-FEB-2025. THE EVENT OCCURRED WITHIN THREE HOURS OF INSERTION. THE INSERTION SITE WAS ABDOMEN. THE INFUSION SET WAS IN USE FOR ONE DAY. THE PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN SUCCESSFULLY. COMPANY DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1567231 | AUTOSOFT 90 | UNO INSET II 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL UM-D | 1002817 | 6000764 | 05705244018129 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Male |