DIMENSION(R) CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2011-00117
- Event Type
- Injury
- Date Received
- July 15, 2011
- Date of Event
- June 11, 2011
- Report Date
- June 22, 2011
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC
- Product Code
- MLM
- PMA / PMN Number
- K060502
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED TACROLIMUS RESULT IS AN INVALID CALIBRATION. USER ERROR, I.E. ACCEPTANCE OF OUTLIERS DURING CALIBRATION, CAUSED THE FALSELY DEPRESSED TACROLIMUS RESULT. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A FALSELY DEPRESSED TACROLIMUS RESULT WAS REPORTED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN. THE SAMPLE WAS LATER REPEATED WITH A NEW REAGENT LOT AND CALIBRATION ON THE SAME INSTRUMENT AND A HIGHER RESULT WAS OBTAINED. A HIGHER RESULT WAS OBTAINED WITH AN ALTERNATE METHODOLOGY. PATIENT DRUG DOSAGE WAS REGULATED ON THE BASIS OF THE DEPRESSED TACROLIMUS RESULT. THERE WAS A REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED TACROLIMUS RESULT. THE CUSTOMER REPORTED PATIENT CONDITION DETERIORATION INCLUDING DEVELOPMENT OF ENCEPHALOPATHY, LOST KIDNEY, DIALYSIS AND VOMITING OF BLOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION(R) CLINICAL CHEMISTRY SYSTEM | TACROLIMUS (TACR) FLEX® REAGENT CARTRIDGE | MLM | SIEMENS HEALTHCARE DIAGNOSTICS INC | FA1284 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | Required Intervention |