FDA Adverse Event Other Summary report: N

ACRYSOF RESTOR

MDR report key: 2163661 · Received June 10, 2011

Report

Report Number
1119421-2011-00713
Event Type
Other
Date Received
June 10, 2011
Date of Event
March 1, 2011
Report Date
May 12, 2011
Manufacturer
ALCON RESEARCH, LTD. / HUNTINGTON
Product Code
MFK
PMA / PMN Number
P040020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION: THE PRODUCT WAS NOT RETURNED FOR ANALYSIS. RESULTS FROM THE PRODUCT HISTORY RECORD REVIEW INDICATED THE PRODUCT MET RELEASE CRITERIA. ROOT CAUSE: NO ROOT CAUSE COULD BE IDENTIFIED BY THE INVESTIGATION. IN THE SURGEON'S OPINION, IT IS UNK AS TO WHETHER THE IOL CAUSED/CONTRIBUTED TO THE OUTCOME. PT IS VERY HAPPY WITH OUTCOME. ACTION TAKEN: INVESTIGATION HAS BEEN COMPLETED BASED ON CURRENT INFORMATION. CONCLUSION: BASED ON OUR CURRENT TRACKING, THERE ARE NO ADVERSE TRENDS FOR THIS REPORTED COMPLAINT AND BASED ON THESE FINDINGS THE RESIDUAL RISK REMAINS UNCHANGED AND AT AN ACCEPTABLE LEVEL. THERE HAVE BEEN NO OTHER SIMILAR COMPLAINTS REPORTED IN THE LOT NUMBER. ADDITIONAL INFORMATION WAS REQUESTED ON 05/13/2011, 05/16/2011, 05/23/2011, AND 06/02/2011 BY PHONE, FAX, AND MAIL. A COMPLETED QUESTIONAIRE HAS NOT BEEN RECEIVED. (B)(4).

Description of Event or Problem · 1

A SURGEON REPORTED A PATIENT WITH A REFRACTIVE SURPRISE FOLLOWING INTRAOCULAR LENS (IOL) IMPLANT SURGERY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACRYSOF RESTOR INTRAOCULAR LENS MFK ALCON RESEARCH, LTD. / HUNTINGTON SN6AD1 12020032

Patients

Seq Age Sex Outcome Treatment
1 Other