BELLATEK HYBRID BAR 5 IMPLANTS
Report
- Report Number
- 0001038806-2025-00619
- Event Type
- Malfunction
- Date Received
- March 18, 2025
- Report Date
- July 10, 2025
- Manufacturer
- BIOMET 3I
- Product Code
- NHA
- UDI-DI
- 00844868001683
- PMA / PMN Number
- K080864
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). A2: AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED. A3: PATIENT SEX UNKNOWN / NOT PROVIDED. A4: WEIGHT UNKNOWN / NOT PROVIDED. B3: DATE OF EVENT UNKNOWN / NOT PROVIDED. E1: LAST/GIVEN NAME UNKNOWN / NOT PROVIDED. G4: ADDITIONAL PMA/510(K) NUMBER ¿ K241032. CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMVIE FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODES WERE ADDED: 4109, 4111 AND 4114. H6: INVESTIGATION FINDINGS CODE WAS ADDED: 213. H6: INVESTIGATION CONCLUSIONS CODE WAS ADDED: 4315. H10: NARRATIVE/DATA WAS UPDATED. ZIMVIE DID NOT RECEIVE THE REPORTED BAR FOR EVALUATION VISUAL INSPECTION COULD NOT BE PERFORMED. THE INVESTIGATION HAS BEEN PERFORMED BASED ON THE AVAILABLE INFORMATION USING THE ITEM NUMBER, DEVICE HISTORY RECORD (DHR) REVIEW, AND RISK MANAGEMENT FILE (RMF). DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. A LOT SPECIFIC COMPLAINT HISTORY REVIEW IS NOT CONDUCTED FOR PSP COMPLAINTS. PATIENT SPECIFIC PRODUCTS HAVE A UNIQUE ONE TIME USE LOT NUMBER. THEREFORE, A LOT SPECIFIC HISTORY SEARCH FOR THE ABUTMENT WAS NOT PERFORMED. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION COULD NOT BE VERIFIED. WITHOUT DEVICE RECEIPT, THE REPORTED EVENT IS NON-VERIFIABLE. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.
IT WAS REPORTED THAT THE BAR BROKE IN HALF IN THE PATIENT¿S MOUTH.
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 599228 | BELLATEK HYBRID BAR 5 IMPLANTS | DENTAL BAR | NHA | BIOMET 3I | 8953460-1 | 00844868001683 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |