FDA Adverse Event Malfunction Summary report: N

BELLATEK HYBRID BAR 5 IMPLANTS

MDR report key: 21634907 · Received March 18, 2025

Report

Report Number
0001038806-2025-00619
Event Type
Malfunction
Date Received
March 18, 2025
Report Date
July 10, 2025
Manufacturer
BIOMET 3I
Product Code
NHA
UDI-DI
00844868001683
PMA / PMN Number
K080864
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). A2: AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED. A3: PATIENT SEX UNKNOWN / NOT PROVIDED. A4: WEIGHT UNKNOWN / NOT PROVIDED. B3: DATE OF EVENT UNKNOWN / NOT PROVIDED. E1: LAST/GIVEN NAME UNKNOWN / NOT PROVIDED. G4: ADDITIONAL PMA/510(K) NUMBER ¿ K241032. CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMVIE FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODES WERE ADDED: 4109, 4111 AND 4114. H6: INVESTIGATION FINDINGS CODE WAS ADDED: 213. H6: INVESTIGATION CONCLUSIONS CODE WAS ADDED: 4315. H10: NARRATIVE/DATA WAS UPDATED. ZIMVIE DID NOT RECEIVE THE REPORTED BAR FOR EVALUATION VISUAL INSPECTION COULD NOT BE PERFORMED. THE INVESTIGATION HAS BEEN PERFORMED BASED ON THE AVAILABLE INFORMATION USING THE ITEM NUMBER, DEVICE HISTORY RECORD (DHR) REVIEW, AND RISK MANAGEMENT FILE (RMF). DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. A LOT SPECIFIC COMPLAINT HISTORY REVIEW IS NOT CONDUCTED FOR PSP COMPLAINTS. PATIENT SPECIFIC PRODUCTS HAVE A UNIQUE ONE TIME USE LOT NUMBER. THEREFORE, A LOT SPECIFIC HISTORY SEARCH FOR THE ABUTMENT WAS NOT PERFORMED. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION COULD NOT BE VERIFIED. WITHOUT DEVICE RECEIPT, THE REPORTED EVENT IS NON-VERIFIABLE. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BAR BROKE IN HALF IN THE PATIENT¿S MOUTH.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
599228 BELLATEK HYBRID BAR 5 IMPLANTS DENTAL BAR NHA BIOMET 3I 8953460-1 00844868001683

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown