EASYTRAK 2
Report
- Report Number
- 2124215-2011-07337
- Event Type
- Injury
- Date Received
- July 15, 2011
- Date of Event
- April 21, 2011
- Report Date
- September 2, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
UPON RECEIPT AT THE POST MARKET QUALITY ASSURANCE LABORATORY THE COMPLETE LEAD WAS RECEIVED. THERE WAS DRIED BLOOD/BODY FLUID IN THE LEAD LUMEN. THE CONDUCTOR COILS WERE DEFORMED 102-107 MILLIMETERS FROM THE TERMINAL PIN, WHICH WAS LIKELY DUE TO A GRABBING TOOL. THERE WERE ALSO CUTS NOTED IN THE INSULATION. THE CLINICAL OBSERVATION WAS UNABLE TO BE REPRODUCED DURING ANALYSIS. THE LEAD WAS FOUND TO BE ELECTRONICALLY CONTINUOUS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD EXHIBITED DISLODGEMENT. THE PATIENT UNDERWENT A REVISION PROCEDURE IN WHICH THIS LEAD WAS EXPLANTED AND REPLACED WITH A COMPETITOR'S PRODUCT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EASYTRAK 2 | IMPLANTABLE LEAD | OJX | CPI - DEL CARIBE | 4543 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Required Intervention | N119| 4543| (B)(4)| 0181| 0171 |