FDA Adverse Event
Malfunction
Summary report: N
BOSTON SCIENTIFIC TRUETOME 44-30MM CUT WIRE 4.4F X 30MM
MDR report key: 21633966
·
Received March 18, 2025
Report
- Report Number
- MW5167766
- Event Type
- Malfunction
- Date Received
- March 18, 2025
- Date of Event
- March 12, 2025
- Report Date
- March 13, 2025
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- NLR
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- NV, US
- Reporter Occupation
- RISK MANAGER
- Health Professional
- Yes
Narratives
Description of Event or Problem · 0
PATIENT UNDERGOING ERCP (ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY). EQUIPMENT MALFUNCTION INTRAOPERATIVELY. EQUIPMENT WAS SWITCHED OUT. NO HARM TO THE PATIENT. TRUETOME 44 30MM CUTE WIRE DID NOT TRIGGER TO CUT DURING PROCEDURE WHEN HOOKED UP TO ERBE MACHINE. NEW ERBE MACHINE WAS BROUGHT INTO PROCEDURE ROOM, TOME WAS PLUGGED IN AND STILL DID NOT FIRE. BOSTON SCIENTIFIC REP STATED IT COULD HAVE POSSIBLY BEEN THE TOME WE WERE USING, SO WE IMMEDIATELY REMOVED FROM PATIENT, SWAPPED IT OUT FOR A NEW TOME AND IT SUCCESSFULLY FIRED. BOSTON SCIENTIFIC TRUETOME 44-30MM CUT WIRE 4.4F X 30MM, GTIN: (B)(4), REF: M00584180, LOT: 34902720.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1211756 | BOSTON SCIENTIFIC TRUETOME 44-30MM CUT WIRE 4.4F X 30MM | UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) | NLR | BOSTON SCIENTIFIC CORPORATION | M00584180 | 34902720 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Female |