CONTAK RENEWAL

MDR report key: 2163316 · Received July 15, 2011

Report

Report Number
2124215-2011-07200
Event Type
Malfunction
Date Received
July 15, 2011
Date of Event
April 21, 2011
Report Date
May 31, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS INVESTIGATION WILL BE UPDATED SHOULD WHEN ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS ELECTIVELY EXPLANTED DUE TO NORMAL BATTERY DEPLETION. THERE WERE NO KNOWN ALLEGATIONS OR COMPLAINTS AGAINST THE DEVICE'S OPERATION, FUNCTIONALITY OR LONGEVITY. UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, INITIAL INVESTIGATION REVEALED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. DETAILED ANALYSIS IS PENDING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H170

Patients

Seq Age Sex Outcome Treatment
1 64 YR N119| H170| 4543| 0148| 1861| 4470