FDA Adverse Event Malfunction Summary report: N

THERABATH PARAFFIN THERAPY BATH

MDR report key: 2163292 · Received July 11, 2011

Report

Report Number
2163292
Event Type
Malfunction
Date Received
July 11, 2011
Date of Event
March 2, 2011
Report Date
July 11, 2011
Manufacturer
WR MEDICAL ELECTRONICS
Product Code
IMC
Product Problem
Yes
Report Source
User Facility report
Reporter Location
SD, US

Narratives

Description of Event or Problem · 1

A PATIENT WAS USING THE PARAFFIN WAX MACHINE AND DEVELOPED A BURN ON DISTAL TIP OF MIDDLE FINGER OF RIGHT HAND. THE THERAPIST CHECKED THE BOTTOM ON THE INSIDE OF THE MACHINE AND IDENTIFIED THAT THERE WERE SEVERAL HOT SPOTS ON THE EDGES ALONG THE BASE (THE WAX WAS NOT TOO HOT). THE BURN WAS A VERY SMALL BLISTER WHICH DID NOT REQUIRE ANY TREATMENT. TEN DAYS AFTER THE OCCURRENCE, A FOLLOW UP CALL ELICITED THE INFORMATION FROM THE PATIENT THAT ALL WAS FINE AND HIS FINGER WAS HEALED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERABATH PARAFFIN THERAPY BATH PARAFFIN WAX BATH IMC WR MEDICAL ELECTRONICS 50 - 60 HZ MODEL TB5 *

Patients

Seq Age Sex Outcome Treatment
1 75 YR