FDA Adverse Event Malfunction Summary report: N

GYNECARE X-TRACT MORCELLATOR

MDR report key: 2163119 · Received July 15, 2011

Report

Report Number
2210968-2011-00899
Event Type
Malfunction
Date Received
July 15, 2011
Date of Event
June 20, 2011
Report Date
June 20, 2011
Manufacturer
ETHICON
Product Code
HET
PMA / PMN Number
K993801
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE REPORTED SYMPTOM OF THE DEVICE NOT ACTIVATING THE DISPOSABLE WAS DUPLICATED. THE CAUSE WAS DUE TO A LOOSE SET SCREW ON THE CPC COUPLER. (B)(4) - DAMAGE TO DRIVE CONNECTOR OR COUPLING.

Additional Manufacturer Narrative · 1

(B)(4). A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED AND THE DEVICE MET ALL FINISHED GOODS RELEASE CRITERIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY ON (B)(6) 2011. DURING THE PROCEDURE, THE MOTOR DRIVE UNIT SPUN THE DEVICE BLADES INTERMITTENTLY AND WAS PULSATING. THIS OCCURRED WITH TWO HANDPIECES. THE PROCEDURE WAS COMPLETED USING A POUCH WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GYNECARE X-TRACT MORCELLATOR LAPAROSCOPIC MORCELLATOR HET ETHICON NA NA

Patients

Seq Age Sex Outcome Treatment
1