FDA Adverse Event Malfunction Summary report: N

ESPRIT VENTILATOR

MDR report key: 2163117 · Received July 15, 2011

Report

Report Number
2031642-2011-00227
Event Type
Malfunction
Date Received
July 15, 2011
Date of Event
June 15, 2011
Report Date
June 15, 2011
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K981072
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE UNIT WOULD NOT POWER ON. THE VENTILATOR WAS NOT IN USE, THEREFORE THERE WAS NO PATIENT INVOLVEMENT OR HARM. THE MANUFACTURER'S SERVICE TECHNICIAN CONFIRMED THE REPORTED PROBLEM. THE SERVICE TECHNICIAN REPLACED THE POWER SUPPLY TO ADDRESS THE FINDING. APPLICABLE FINAL TESTING WAS COMPLETED AND TESTS PASSED TO OPERATING SPECIFICATIONS. ANALYSIS OF THE POWER SUPPLY REVEALED A MALFUNCTION THAT MAY RESULT IN THE UNIT SHUTTING DOWN DURING OPERATION WITH AN ACTIVE BACKUP ALARM WHEN AC POWER IS RESTORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA