FDA Adverse Event Injury Summary report: N

MINICAP TRANSFER SET

MDR report key: 2163098 · Received July 15, 2011

Report

Report Number
1423500-2011-09058
Event Type
Injury
Date Received
July 15, 2011
Date of Event
June 1, 2011
Report Date
June 20, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ROOT CAUSE OF THE REPORTED CONDITION OF PERITONITIS IS USE ERROR - POOR ASEPTIC TECHNIQUE. A BATCH REVIEW OF THE POTENTIALLY ASSOCIATED LOT NUMBER (H10H31055) REVEALED NO EXCEPTIONS DURING THE MANUFACTURING PROCESS. A LABELING REVIEW WAS PERFORMED AND THE LABELING WAS FOUND TO PROVIDE AMPLE INSTRUCTIONS RELATED TO PREVENTION OF THE USE ERROR IN ASEPTIC TECHNIQUE. BAXTER HAS RECEIVED SIMILAR REPORTS FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.

Additional Manufacturer Narrative · 1

(B)(4). AS THE DATE OF ONSET OF THIS PERITONITIS EVENT IS UNKNOWN THE SAMPLE WAS NOT REQUESTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, AND/OR UPON CONCLUSION OF BAXTER'S INVESTIGATION A FOLLOW-UP REPORT WILL BE SUBMITTED. THIS IS THE FIRST OF FIVE REPORTS ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 1

BAXTER CONTACTED THE PERITONEAL DIALYSIS NURSE (PDRN) REGARDING A RELATED ISSUE OF THE HP NOT FEELING WELL. THE PDRN STATED THAT THE HP HAD PERITONITIS AND WAS HOSPITALIZED FROM (B)(6) 2011- (B)(6) 2011. CULTURES WERE OBTAINED ON (B)(6) 2011. TREATMENT INITIATED WHILE HOSPITALIZED WAS WITH INTRAPERITONEAL (IP) VANCOMYCIN 3 TIMES A WK AND CEFTAZIDIME DAILY, BOTH 1 GM DOSES. THE HP RECEIVED THE LAST IP DOSE OF ANTIBIOTICS ON (B)(6) 2011. CAUSALITY WAS GIVEN AS AN UNTRAINED CAREGIVER ASSISTING THE HP WITH PD THERAPY. THE HP'S RECOVERY IS ONGOING AND IMPROVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 86 YR Hospitalization| R HOME CHOICE CYCLER| LOWCAL (PD4) AMBUFLEX