FDA Adverse Event Malfunction Summary report: N

INFUSO.R.

MDR report key: 2163060 · Received July 15, 2011

Report

Report Number
6000001-2011-12140
Event Type
Malfunction
Date Received
July 15, 2011
Date of Event
June 1, 2011
Report Date
June 28, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K883577
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS RETURNED FOR EVALUATION HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. ONCE THE EVALUATION IS COMPLETE, A FOLLOW UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED MALFUNCTION OF "PRONGS ARE BROKEN OFF WHERE SYRINGE IS HELD" WAS CONFIRMED AND NOT REPRODUCED. THE ROOT CAUSE WAS ATTRIBUTED TO A DAMAGED SYRINGE HOLDER. THE SYRINGE HOLDER WAS REPLACED TO FIX THE PROBLEM. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED AN INFUSO.R. PUMP WITH A CONDITION OF "PRONGS ARE BROKEN OFF WHERE SYRINGE IS HELD." THIS CONDITION OCCURRED DURING PROGRAMMING/SETUP. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO PATIENT INVOLVEMENT WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSO.R. PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1