FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 2163020
·
Received July 15, 2011
Report
- Report Number
- 2050012-2011-03098
- Event Type
- Malfunction
- Date Received
- July 15, 2011
- Date of Event
- June 12, 2011
- Report Date
- June 12, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND FOUND THAT THE CLOT/OBSTRUCTION DETECTOR (ODC) FOR THE MC PROBE HAD A RUPTURED DIAPHRAGM. THE FSE REPLACED THE ODC ASSEMBLY, THEN CALIBRATED THE ODC, AND TESTED THE CIRCUIT. THE FSE ALSO CALIBRATED AND RAN QC ON THE SYSTEM AND ALL RESULTS PASSED WITHIN SPECIFICATIONS. REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THERE WAS A FLUID LEAK FROM THE MODULAR CARTRIDGE (MC) SAMPLE PROBE TRANSDUCER IN THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE CUSTOMER INDICATED THAT THE LEAK OCCURRED AFTER REPLACING THE SAMPLE PROBE DUE TO AN MC PROBE OBSTRUCTION. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 800 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |