FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600 SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 2163019
·
Received July 15, 2011
Report
- Report Number
- 2050012-2011-03097
- Event Type
- Malfunction
- Date Received
- July 15, 2011
- Date of Event
- June 12, 2011
- Report Date
- June 12, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 AND INSPECTED THE SOURCE OF THE LEAK AND FOUND THE ELECTROLYTE INJECTION CUP (EIC) OVERFLOWING. THE FSE CLEANED THE EIC PORTS AND VALVE SEATS. THE REPAIR WAS VERIFIED PER ESTABLISHED PROCEDURES.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THERE WAS A LEAK COMING FROM THE MODULAR CHEMISTRY (MC) SIDE OF THE UNICEL DXC 600 SYNCHRON CHEMISTRY ANALYZER. CUSTOMER STATED THAT NO ERRORS WERE GENERATED AND ALL CHEMISTRIES WERE OPERATING PROPERLY. THE CUSTOMER ALSO INDICATED THAT THE LIQUID WAS CLEAR AND THAT THE SOURCE OF THE LEAK WAS UNKNOWN. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |