FDA Adverse Event Malfunction Summary report: N

PROXIMATE** RELOADABLE LINEAR STAPLER - 60 MM

MDR report key: 2162967 · Received July 15, 2011

Report

Report Number
3005075853-2011-02893
Event Type
Malfunction
Date Received
July 15, 2011
Date of Event
June 20, 2011
Report Date
June 23, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED ON THE DEVICE. THE RELOAD WAS RECEIVED FULLY LOADED WITH STAPLES. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH THE RETURNED RELOAD AND IT FIRED AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLES WERE NOTED TO HAVE A PROPER B-FORMATION AND THE STAPLE LINE WAS COMPLETE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A COLECTOMY PROCEDURE, THE DEVICE WAS PLACED ACROSS THE BOWEL AND FIRED. ONCE REMOVED, THEY FOUND THE DEVICE HAD CUT BUT NOT STAPLED THE TISSUE. THE SECTION OF THE TISSUE NOT STAPLED WAS REMOVED. A NEW DEVICE AND CARTRIDGE WAS FIRED ACROSS THE AREA SUCCESSFULLY. THE PROCEDURE WAS COMPLETED WITH NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE** RELOADABLE LINEAR STAPLER - 60 MM STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA H43F3N

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE