FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 2162702
·
Received July 14, 2011
Report
- Report Number
- 2050012-2011-02963
- Event Type
- Malfunction
- Date Received
- July 14, 2011
- Date of Event
- June 12, 2011
- Report Date
- June 12, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
LEVEL SENSE WAS NOT WORKING PROPERLY SO BCI CTS (CUSTOMER TECHNICAL SUPPORT) DIRECTED THE CUSTOMER TO MOVE SENSOR CIRCUIT CLOSER TO BOTTLE TO GIVE CORRECT READING. THIS RESOLVED THE ISSUE. NO SERVICE WAS REQUESTED FOR THIS EVENT. THE ISSUE WAS RESOLVED THROUGH TROUBLESHOOTING WITH THE CTS.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REPORTING THAT THE NO FOAM HOSE IN THE UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER CAME LOOSE FROM THE BOTTLE AND CAUSED LEAKING. THE HOSE WAS REATTACHED AND NO FURTHER LEAKING WAS OBSERVED. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |