VITALITY 2
Report
- Report Number
- 2124215-2011-06876
- Event Type
- Malfunction
- Date Received
- July 14, 2011
- Date of Event
- April 19, 2011
- Report Date
- April 28, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE. THE DEVICE WAS THEN SUBJECTED TO, AND PASSED, A SERIES OF DIAGNOSTIC TESTS TO ASSESS THE PERFORMANCE OF DEFIBRILLATION, PACING, SENSING, HIGH VOLTAGE SHOCKING, AND DIAGNOSTIC RECORDING FUNCTIONS.
THE DEVICE IS CURRENTLY UNDERGOING ANALYSIS. WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED AND RETURNED FOR NORMAL BATTERY DEPLETION. NO FIELD ALLEGATIONS HAVE BEEN REPORTED AGAINST THIS PRODUCT. THIS REPORT WAS CREATED DUE TO INITIAL LABORATORY FINDINGS AS THIS PRODUCT DID NOT PASS LONGEVITY CALCULATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T167 |