FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 2162543 · Received July 14, 2011

Report

Report Number
2124215-2011-07793
Event Type
Malfunction
Date Received
July 14, 2011
Date of Event
April 20, 2011
Report Date
April 20, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Removal / Correction Number
Z-1620-09 TO Z-1621-09
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. THERE WAS NO ALLEGATION FROM THE FIELD REGARDING THE FUNCTION OF THE DEVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED IN ASSOCIATION WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H170

Patients

Seq Age Sex Outcome Treatment
1 79 YR H220| 4543| 0184| H170| 4472