FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2162456 · Received July 14, 2011

Report

Report Number
6000001-2011-11920
Event Type
Malfunction
Date Received
July 14, 2011
Date of Event
June 15, 2011
Report Date
June 20, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH AN ERROR CODE 703 WAS NOT CONFIRMED BY BAXTER PERSONNEL DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS ASSIGNED TO ISSUES WITH THE USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB). THE UIM PCB WAS REPLACED TO CORRECT THIS CONDITION. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 6.13.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE HAS NOT BEEN SERVICED PRIOR TO THIS EVENT FOR THE REPORTED CONDITION.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT A COLLEAGUE INFUSION PUMP IN WHICH FAILURE CODE 703 OCCURRED. THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THE EVENT OCCURRED IN THE GENERAL PATIENT WARD. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN.