MITEK VAPR S90 ELECTRODE
Report
- Report Number
- 1221934-2011-00261
- Event Type
- Malfunction
- Date Received
- July 14, 2011
- Date of Event
- June 7, 2011
- Report Date
- June 7, 2011
- Manufacturer
- DEPUY MITEK
- Product Code
- GEI
- PMA / PMN Number
- K041135
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
MITEK IS AT THIS POINT IN TIME IN THE INFORMATION GATHERING MODE. WHEN ALL THAT CAN BE HAD, IS HAD AND THOROUGHLY INVESTIGATED AND EVALUATED, THOSE RESULTS WILL BE THE SUBJECT MATTER IN A FOLLOW-UP REPORT.
MITEK RECEIVED 3 DEVICES: 2 NEW DEVICES STILL IN THEIR UNOPENED PACKAGING AND 1 DEVICE WITH SOME APPARENT DAMAGE: WE ASSUME THAT THE DEVICE WITH THE DAMAGE IS THE COMPLAINT DEVICE. THE TWO "NEW" DEVICES WERE REMOVED FROM THEIR PACKAGING AND THEN EVALUATED BOTH VISUALLY AND MECHANICALLY. FIRST, THE DEVICES WERE VIEWED BOTH WITH THE NAKED EYE AND UNDER POWER; THEY APPEAR IN GOOD NEW CONDITION, NO DAMAGE AND NO ANOMALIES. THE DEVICES WERE THEN MATED WITH A STANDARD VAPR TEST SET UP; THE TEST GENERATOR RECOGNIZED THE ELECTRODES AND THE PROPER DEFAULTS CAME UP ON THE DISPLAY. FURTHER WHEN THE FOOTSWITCH PEDALS WERE ACTIVATED THERE WAS EVIDENCE OF POWER AT THE TIP OF THE ELECTRODES. THIS PROCEDURE WAS REPEATED SEVERAL TIMES TO INSURE CONTINUITY OF FUNCTION; NO FAULT FOUND. THE ASSUMED COMPLAINT DEVICE WAS SUBJECTED TO A VISUAL EXAMINATION AND ALTHOUGH THE COMPLAINT NARRATIVE DESCRIBES THAT SOME OF THE BLACK INSULATION MATERIAL CAME AWAY FROM THE DEVICE, WE COULD SEE NO EVIDENCE OF THIS, HOWEVER, THERE IS DAMAGE TO THE DISTAL TIP OF THE DEVICE, THERE IS MASS MISSING AT THE 7 0'CLOCK POSITION. A BATCH RECORD REVIEW HAS BEEN CONDUCTED TO DETERMINE IF THERE WERE ANY INTERNAL PROCESSING ISSUES WHICH WOULD HAVE CONTRIBUTED TO THE NATURE OF THE PRODUCT COMPLAINT. OUR RESULTS INDICATE THAT THIS BATCH OF PRODUCT WAS PROCESSED WITHOUT INCIDENT AND THEREFORE THERE IS NO INTERNALLY ASSIGNABLE CAUSE FOR THE REPORTED PROBLEM. FURTHER, A REVIEW INTO THE MITEK COMPLAINTS SYSTEM REVEALED NO OTHER SIMILAR COMPLAINT FOR THIS LOT OF DEVICES THAT WERE RELEASED TO DISTRIBUTION. THESE ISSUES WITH THE DISTAL TIP OF S90 AND LPS ELECTRODES HAVE BEEN THE FOCUS OF A LONG AND INTENSIVE INVESTIGATION, WHICH HAS BEEN CONDUCTED IN COLLABORATION BETWEEN MITEK AND THE MANUFACTURER. THE BULK OF THE RETURNED S90 DEVICES WITH REPORTED DISTAL TIP FAILURE MODES: SPARKING AND ARCHING, MISSING MASS, BREAKING OFF, ETC., HAVE BEEN RECEIVED AT MITEK, VISUALLY EVALUATED TO SUBSTANTIATE THE FAILURE MODE, AND THEN WERE FORWARDED TO THE MANUFACTURER TO SUPPORT THEIR ONGOING INVESTIGATION TOWARDS DETERMINING WHAT THE ROOT CAUSE OR UNDERLYING REASON FOR THESE ISSUES COULD BE. TO DATE, THE ANALYSIS OF COMPLAINT DEVICES FOR THIS FAILURE MODE HAS NOT BEEN ABLE TO DETERMINE A SPECIFIC ROOT CAUSE FOR EACH OF THE DEVICE FAILURES, AND HAS ALSO NOT BEEN ABLE TO IDENTIFY A GENERIC ROOT CAUSE TO EXPLAIN THE FAILURES AS A WHOLE. PRODUCT PROBLEM INVESTIGATIONS AND CORRECTIVE ACTION ACTIVITY HAVE NOT YIELDED ANY FAILURE REASONS THAT ARE OUTSIDE OF THE HISTORICAL HYPOTHESIS THAT TECHNIQUE IS THE MOST LIKELY DRIVING FACTOR: TIP BURIAL ACTIVATION: THIS CAN CAUSE CARBON TRACKING BETWEEN ACTIVE AND RETURN CREATING AN "OUTPUT SHORT" ERROR CODE ON THE GENERATOR; THIS IS CONSIDERED TO BE A USER ISSUE AND EXTERNAL FAILURE. ACTIVATION OUTSIDE OF THE SALINE FIELD: THIS CAN CAUSE THERMAL DAMAGE TO THE TIP, IN TURN REDUCING THE DISTANCE BETWEEN THE ACTIVE AND RETURN PATHS; ALSO, THE FAILURE MODE AND FREQUENCY RATE FOR THIS DEVICE ARE WELL WITHIN THE FMEA RISK ANALYSIS. BEYOND OUR HYPOTHESIS AND CONSIDERATION, WE CANNOT DISCERN ANY OTHER ROOT CAUSE FOR THESE ISSUES. OUTSIDE OF CONTINUED ANALYSIS ACTIVITY AND TRENDING, NO FURTHER ACTION IS WARRANTED AT THIS TIME, HOWEVER, IF AND WHEN A DEFINITIVE ROOT CAUSE CAN BE DETERMINED, IT WILL BE NOTED IN ANY FUTURE RELATIVE EVALUATION SUMMARIES.
OUR AFFILIATE IS REPORTING TO US THAT DURING AN ARTHROSCOPIC KNEE REPAIR, A PORTION OF THE DISTAL TIP, OR SOME OF THE INSULATION COATING AT THE DISTAL TIP CAME OFF INTO THE PATIENT & APOS'S JOINT SPACE. IT IS NOT KNOWN IF ALL OF THE FRAGMENTS/DEBRIS WAS REMOVED FROM THE JOINT SPACE; HOWEVER, THE PROCEDURE WAS CONCLUDED SUCCESSFULLY WITHOUT FURTHER ISSUE OR HARM TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MITEK VAPR S90 ELECTRODE | ELECTROSURGICAL, CUTTING & COAGULATING | GEI | DEPUY MITEK | NA | M1101031 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |