HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-09031
- Event Type
- Malfunction
- Date Received
- July 14, 2011
- Date of Event
- June 20, 2011
- Report Date
- June 20, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS INCIDENT WAS DETERMINED TO BE CAUSED BY USE/USER ERROR AND THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION. THEREFORE A BATCH REVIEW AND SAMPLE EVALUATION WILL NOT BE CONDUCTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SENT.
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 WAS NOT CONFIRMED; HOWEVER, PER THE COMPLAINT INFORMATION THE CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE PATIENT DISCONNECTED THE PATIENT LINE FROM THE TRANSFER SET. THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. THIS REVIEW FINDS THE LABELING ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.
(B)(4). THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4).
A HOME PATIENT (HP) CONTACTED (B)(4) REGARDING A SYSTEM ERROR (SE) 2240 ALARM THAT OCCURRED ON THE HOME CHOICE (HC) UNIT DURING DWELL. THE HP STATES SHE DISCONNECTED TO USE RESTROOM SHE DID NOT PLACE A CAP ON PATIENT LINE THEN RECONNECTED. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP WITH CLEARING THE ALARM THEN EXPLAINED A SE 2240 INDICATES A LARGE AMOUNT OF AIR HAS ENTERED SETUP AND THAT THE HP NEEDED TO END THERAPY AND CONTACT THEIR NURSE. THERE WAS PATIENT INVOLVEMENT. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED. A FOLLOW UP WAS DONE VIA PHONE CALL. THE REGISTERED NURSE (RN) WAS NOT AWARE OFF THE OCCURRENCE. THE RN WAS MADE AWARE OF THE USE ERROR THAT OCCURRED. THE RN STATED THE HOME PATIENT (HP) HAS BEEN DOING THERAPY FINE. THERE WERE NO FURTHER DETAILS PROVIDED
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | HOME CHOICE |