FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2162355 · Received July 14, 2011

Report

Report Number
2531779-2011-04952
Event Type
Malfunction
Date Received
July 14, 2011
Report Date
April 28, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION COMPLETED ON 06/17/2011. (B)(6). RECALL 2531779-03/24/2010-003-R. DURING EVALUATION THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE PHYSICALLY DAMAGED.

Description of Event or Problem · 1

PUMP IS RETURNED FOR MULTIPLE LOSS OF PRIME ALARMS. EVALUATION REVEALED A DAMAGED FORCE SENSOR ASSEMBLY. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 16 YR