FDA Adverse Event Injury Summary report: N

RX BILIARY

MDR report key: 2162186 · Received July 14, 2011

Report

Report Number
3005099803-2011-02302
Event Type
Injury
Date Received
July 14, 2011
Date of Event
June 10, 2011
Report Date
June 20, 2011
Manufacturer
BOSTON SCIENTIFIC - SPENCER
Product Code
FGE
PMA / PMN Number
K965147
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED; HOWEVER THE EVALUATION HAS NOT YET BEEN COMPLETED. IF THERE IS ANY FURTHER RELEVANT INFORMATION, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: A VISUAL EVALUATION OF THE RETURNED DEVICE REVEALED THAT THE GUIDE CATHETER WAS DETACHED/SEPARATED FROM THE PULL WIRE AND WAS KINKED AT MULTIPLE LOCATIONS. THE WORKING LENGTH OF GUIDE CATHETER WAS FOUND INTACT (UNBROKEN). THE PROXIMAL END OF THE GUIDE CATHETER THAT ATTACHES TO THE WELDED CANNULA/PULL WIRE ASSEMBLY WAS FREE OF SPLITS/TEARS. THE OVERALL LENGTH OF THE GUIDE CATHETER MET SPECIFICATION OF 32.0 CM (MINIMUM). THE PULL WIRE WAS BENT AT THE PROXIMAL END. THE WHITE MOLDED HANDLE WAS DETACHED FROM THE PULL WIRE. THE PROXIMAL TIP OF THE PULL WIRE WAS BENT WHICH INDICATED THAT THE MOLDED HANDLE WAS SECURELY ATTACHED DURING MANUFACTURING. THE WELDED CANNULA/PULL WIRE WAS FREE OF DAMAGE. BASED ON THE CONDITION OF THE DEVICE AND THE DETAILS OF THE EVENT, THE MOST PROBABLE ROOT CAUSE FOR THIS COMPLAINT IS OPERATIONAL CONTEXT. DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED LOT.

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION RECEIVED SINCE (B)(6), 2011: THE OVERALL LENGTH OF THE GUIDE CATHETER MEASURED APPROXIMATELY 32.8 CM WHICH DOES NOT MEET THE GUIDE CATHETER CUT LENGTH SPECIFICATION OF 32.0CM +/- 1MM. THE INCREASED LENGTH OF THE GUIDE CATHETER IS DUE TO KINKS IN ITS WORKING LENGTH; THE KINKS FLATTEN THE GUIDE CATHETER ASSEMBLY CAUSING THE OVERALL INTEGRITY TO BE COMPROMISED. IT WAS ORIGINALLY REPORTED THAT THE OVERALL LENGTH OF THE GUIDE CATHETER MET SPECIFICATION.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA RX BILIARY STENT SYSTEM WAS USED DURING AN ERCP (ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY) PROCEDURE ABOVE BIFURCATION OF THE HEPATIC DUCT OF A (B)(6) OLD MALE PATIENT WEIGHING (B)(6). DURING THE PROCEDURE, AS THE PULLWIRE WAS RETRACTED FOR STENT DEPLOYMENT, THE PULLWIRE BROKE IN HALF AND THE GUIDE CATHETER DETACHED FROM DELIVERY SYSTEM. THE PULLWIRE WAS THEN PULLED ALL THE WAY OUT OF THE CATHETER. THE DETACHED GUIDE CATHETER REMAINED IN THE DEPLOYED STENT INSIDE THE PATIENT. THE GUIDE CATHETER WAS RETRACTED FROM THE PATIENT WITH A SNARE AND THE STENT WAS LEFT INSIDE THE PATIENT AT THE TARGET SITE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO REPORTED PATIENT COMPLICATIONS. THE PATIENT WAS REPORTED TO BE FINE AFTER THE PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA RX BILIARY STENT SYSTEM WAS USED DURING AN ERCP (ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY) PROCEDURE ABOVE BIFURCATION OF THE HEPATIC DUCT OF A (B)(6) OLD MALE PATIENT WEIGHING (B)(6). DURING THE PROCEDURE, AS THE PULLWIRE WAS RETRACTED FOR STENT DEPLOYMENT, THE PULLWIRE BROKE IN HALF AND THE GUIDE CATHETER DETACHED FROM DELIVERY SYSTEM. THE PULLWIRE WAS THEN PULLED ALL THE WAY OUT OF THE CATHETER. THE DETACHED GUIDE CATHETER REMAINED IN THE DEPLOYED STENT INSIDE THE PATIENT. THE GUIDE CATHETER WAS RETRACTED FROM THE PATIENT WITH A SNARE AND THE STENT WAS LEFT INSIDE THE PATIENT AT THE TARGET SITE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO REPORTED PATIENT COMPLICATIONS. THE PATIENT WAS REPORTED TO BE FINE AFTER THE PROCEDURE. ON (B)(6), 2011, THE COMPLAINANT CLARIFIED THE PULLWIRE DID NOT BREAK IN HALF.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA RX BILIARY STENT SYSTEM WAS USED DURING AN ERCP (ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY) PROCEDURE ABOVE BIFURCATION OF THE HEPATIC DUCT OF A (B)(6) MALE PATIENT WEIGHING (B)(6). DURING THE PROCEDURE, AS THE PULLWIRE WAS RETRACTED FOR STENT DEPLOYMENT, THE PULLWIRE BROKE IN HALF AND THE GUIDE CATHETER DETACHED FROM DELIVERY SYSTEM. THE PULLWIRE WAS THEN PULLED ALL THE WAY OUT OF THE CATHETER. THE DETACHED GUIDE CATHETER REMAINED IN THE DEPLOYED STENT INSIDE THE PATIENT. THE GUIDE CATHETER WAS RETRACTED FROM THE PATIENT WITH A SNARE AND THE STENT WAS LEFT INSIDE THE PATIENT AT THE TARGET SITE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO REPORTED PATIENT COMPLICATIONS. THE PATIENT WAS REPORTED TO BE FINE AFTER THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RX BILIARY CATHETER, BILIARY, DIAGNOSTIC FGE BOSTON SCIENTIFIC - SPENCER M00545590 14160672

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention