FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2162173 · Received July 7, 2011

Report

Report Number
3004209178-2011-82098
Event Type
Injury
Date Received
July 7, 2011
Date of Event
June 21, 2011
Report Date
June 25, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED FOR DIABETIC KETOACIDOSIS. THE BLOOD GLUCOSE READING WAS 868 MG/DL. IT WAS STATED THAT THE CUSTOMER WAS FEELING ILL AND WAS EXPERIENCING UNEXPLAINED HIGH BLOOD GLUCOSE. IT WAS STATED THAT THE SPOUSE REQUESTED TESTING THE INSULIN PUMP BEFORE THE CUSTOMER WEARS IT AGAIN. TROUBLESHOOTING WAS PERFORMED. THE PROGRAMMING, TIME, AND DATE WERE CORRECT. RAN A FIXED PRIME AND HIGH PRESSURE TEST AND THE INSULIN PUMP PASSED THE TESTS. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-522NAS

Patients

Seq Age Sex Outcome Treatment
1 50 YR Hospitalization PARADIGM, 6MM CATHETER| QUICK-SET, 23| UNOMEDICAL INFUSION SET, MMT-399