ENDURANT STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2011-01256
- Event Type
- Injury
- Date Received
- July 8, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 8, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P100021
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION, RESULTS: ENDOLEAK, BILATERAL ILIAC ARTERY THROMBUS. EVALUATION, CONCLUSION: BILATERAL ILIAC ARTERY THROMBUS.
AN ENDURANT BIFURCATED STENT GRAFT WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF BILATERAL ILIAC ARTERY ANEURYSMS WITH THROMBUS, ON AN UNK DATE. AT THE TIME OF IMPLANT, THE LEFT HYPOGASTRIC ARTERY WAS OCCLUDED WITH COILS. A ROUTINE CT, ONE MONTH AGO, SHOWED THAT THE RIGHT ILIAC PORTION OF THE STENT GRAFT HAD LOST ITS SEAL, RESULTING IN A DISTAL TYPE I ENDOLEAK, AS THE THROMBUS WAS TOO SOFT TO MAINTAIN THE SEAL. APPROXIMATELY 4 WEEKS AGO, A 20X12 ANEURX BIFURCATED STENT GRAFT WAS PLACED VIA THE RIGHT SIDE. THE CONTRALATERAL GATE WAS CANNULATED VIA BRACHIAL ACCESS AND PLACED INTO THE RIGHT HYPOGASTRIC SUCCESSFULLY, FOLLOWED BY THE PLACEMENT OF ANOTHER MANUFACTURER'S STENT. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDURANT STENT GRAFT SYSTEM | MIH | MEDTRONIC CARDIOVASCULAR IRELAND | NA | V00831427 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Required Intervention |