FDA Adverse Event Injury Summary report: N

ENDURANT STENT GRAFT SYSTEM

MDR report key: 2161864 · Received July 8, 2011

Report

Report Number
2953200-2011-01256
Event Type
Injury
Date Received
July 8, 2011
Date of Event
June 1, 2011
Report Date
June 8, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR IRELAND
Product Code
MIH
PMA / PMN Number
P100021
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION, RESULTS: ENDOLEAK, BILATERAL ILIAC ARTERY THROMBUS. EVALUATION, CONCLUSION: BILATERAL ILIAC ARTERY THROMBUS.

Description of Event or Problem · 1

AN ENDURANT BIFURCATED STENT GRAFT WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF BILATERAL ILIAC ARTERY ANEURYSMS WITH THROMBUS, ON AN UNK DATE. AT THE TIME OF IMPLANT, THE LEFT HYPOGASTRIC ARTERY WAS OCCLUDED WITH COILS. A ROUTINE CT, ONE MONTH AGO, SHOWED THAT THE RIGHT ILIAC PORTION OF THE STENT GRAFT HAD LOST ITS SEAL, RESULTING IN A DISTAL TYPE I ENDOLEAK, AS THE THROMBUS WAS TOO SOFT TO MAINTAIN THE SEAL. APPROXIMATELY 4 WEEKS AGO, A 20X12 ANEURX BIFURCATED STENT GRAFT WAS PLACED VIA THE RIGHT SIDE. THE CONTRALATERAL GATE WAS CANNULATED VIA BRACHIAL ACCESS AND PLACED INTO THE RIGHT HYPOGASTRIC SUCCESSFULLY, FOLLOWED BY THE PLACEMENT OF ANOTHER MANUFACTURER'S STENT. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDURANT STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR IRELAND NA V00831427

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention