FDA Adverse Event Injury Summary report: N

TALENT THORACIC STENT GRAFT WITH XCELERANT

MDR report key: 2161863 · Received July 8, 2011

Report

Report Number
2953200-2011-01257
Event Type
Injury
Date Received
July 8, 2011
Date of Event
June 1, 2011
Report Date
June 8, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR IRELAND
Product Code
MIH
PMA / PMN Number
P070007
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION, RESULTS: ENDOLEAK, DISEASE PROGRESSION AND/OR DILATATION OF THE DISTAL AORTIC NECK. EVALUATION, CONCLUSION: DISEASE PROGRESSION AND/OR DILATATION OF THE DISTAL AORTIC NECK.

Description of Event or Problem · 1

THREE TALENT THORACIC STENT GRAFTS WERE IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN AORTIC ANEURYSM, APPROXIMATELY SEVEN MONTHS AGO WITHOUT ISSUES. A ROUTINE CT LAST MONTH SHOWED THAT A DISTAL TYPE I ENDOLEAK WITH POSSIBLE ANEURYSMAL DISEASE PROGRESSION AND/OR DILATATION OF THE DISTAL AORTIC NECK. AN INTERVENTION OCCURRED ONE MONTH AGO WITH TWO ADDITIONAL STENT GRAFTS; A TALENT DISTAL MAIN STENT GRAFT WAS PLACED DISTALLY; AND A TALENT CAPTIVIA STENT GRAFT WAS PLACED TO OVERLAP THE NEW TALENT DISTAL MAIN STENT GRAFT AND THE ORIGINAL TALENT DISTAL STENT GRAFT. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT THORACIC STENT GRAFT WITH XCELERANT MIH MEDTRONIC CARDIOVASCULAR IRELAND NA V00480645

Patients

Seq Age Sex Outcome Treatment
1 83 YR Required Intervention