FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2161862 · Received July 14, 2011

Report

Report Number
2531779-2011-04939
Event Type
Malfunction
Date Received
July 14, 2011
Date of Event
April 5, 2011
Report Date
April 5, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION COMPLETED ON 06/15/2011. (B)(6). 2531779-03/24/2010-003-R. DURING EVALUATION THE FORCE SENSOR WAS FOUND TO HAVE AN INTERMITTENT CONNECTION AT THE FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY SHOWED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURED WHICH WAS NOT DUPLICATED DURING TESTING.

Description of Event or Problem · 1

DURING EVALUATION, AN INTERMITTENT CONNECTION WAS FOUND AT THE FORCE SENSOR PINS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 67